2002Journal of Clinical MicrobiologyOpen access

Accuracy of Results Obtained by Performing a Second Ligase Chain Reaction Assay and PCR Analysis on Urine Samples with Positive or Near-Cutoff Results in the LCx Test for Chlamydia trachomatis

Santina Castriciano, K. Luinstra, D. Jang, Jay S. Patel, James B. Mahony, J. Kapala, Max Chernesky

Open full text 27 citations

Abstract

Nucleic acid amplification assays such as the ligase chain reaction and PCR have encountered reproducibility problems. The initial extract and a newly extracted aliquot of urine specimens (n = 120) which had signal-to-cutoff (S/CO) ratios above 0.80 by the LCx Chlamydia assay were retested. Nucleic acid was extracted from an additional urine sample for testing by the AMPLICOR PCR Chlamydia assay. Fifteen percent (18 of 120) of the urine specimens were negative by all repeat tests (initial mean S/CO ratio by the LCx Chlamydia assay, 0.93; S/CO ratio range, 0.80 to 3.30). Repeat testing of the 102 specimens with possible positive results by the LCx Chlamydia assay by use of the initially extracted aliquot confirmed the results for 95 (93.1%) of the specimens; repeat testing of a newly extracted aliquot confirmed the results for 87 (85.3%) of the specimens. Twenty specimens had discordant results by the two repeat LCx Chlamydia assays. A total of 78 of 102 (76.5%) of the specimens were positive by the AMPLICOR PCR, and the AMPLICOR PCR confirmed the results for 82.1% (78 of 95) and 89.6% (78 of 87) of the specimens positive by the two repeat LCx Chlamydia assays, respectively. Some of the discrepancies observed by multiple repeat tests may have been due to specimen mislabeling or contamination during performance of the procedure rather than to the LCx Chlamydia assay. Both assays suffered from a lack of reproducibility on repeat testing with a small proportion of specimens, probably due to the presence of low levels of DNA, the presence of variable amounts of amplification inhibitors, and the loss of DNA during extraction.

About this research paper

What this paper is about

Nucleic acid amplification assays such as the ligase chain reaction and PCR have encountered reproducibility problems. The initial extract and a newly extracted aliquot of urine specimens (n = 120) which had signal-to-cutoff (S/CO) ratios above 0.80 by the LCx Chlamydia assay were retested. Nucleic acid was extracted from an additional urine sample for testing by the AMPLICOR PCR Chlamydia assay. Fifteen percent (18 of 120) of the urine specimens were negative by all repeat tests (initial mean S/CO ratio by the LCx Chlamydia assay, 0.93; S/CO ratio range, 0.80 to 3.30). Repeat testing of the 102 specimens with possible positive results by the LCx Chlamydia assay by use of the initially extracted aliquot confirmed the results for 95 (93.1%) of the specimens; repeat testing of a newly extracted aliquot confirmed the results for 87 (85.3%) of the specimens. Twenty specimens had discordant results by the two repeat LCx Chlamydia assays. A total of 78 of 102 (76.5%) of the specimens were positive by the AMPLICOR PCR, and the AMPLICOR PCR confirmed the results for 82.1% (78 of 95) and 89.6% (78 of 87) of the specimens positive by the two repeat LCx Chlamydia assays, respectively. Some of the discrepancies observed by multiple repeat tests may have been due to specimen mislabeling or contamination during performance of the procedure rather than to the LCx Chlamydia assay. Both assays suffered from a lack of reproducibility on repeat testing with a small proportion of specimens, probably due to the presence of low levels of DNA, the presence of variable amounts of amplification inhibitors, and the loss of DNA during extraction.

Why it matters

OpenAlex reports 27 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Nucleic acid amplification assays such as the ligase chain reaction and PCR have encountered reproducibility problems. The initial extract and a newly extracted aliquot of urine specimens (n = 120) which had signal-to-cutoff (S/CO) ratios above 0.80 by the LCx Chlamydia assay were retested. Nucleic acid was extracted from an additional urine sample for testing by the AMPLICOR PCR Chlamydia assay. Fifteen percent (18 of 120) of the urine specimens were negative by all repeat tests (initial mean S/CO ratio by the LCx Chlamydia assay, 0.93; S/CO ratio range, 0.80 to 3.30). Repeat testing of the 102 specimens with possible positive results by the LCx Chlamydia assay by use of the initially extracted aliquot confirmed the results for 95 (93.1%) of the specimens; repeat testing of a newly extracted aliquot confirmed the results for 87 (85.3%) of the specimens. Twenty specimens had discordant results by the two repeat LCx Chlamydia assays. A total of 78 of 102 (76.5%) of the specimens were positive by the AMPLICOR PCR, and the AMPLICOR PCR confirmed the results for 82.1% (78 of 95) and 89.6% (78 of 87) of the specimens positive by the two repeat LCx Chlamydia assays, respectively. Some of the discrepancies observed by multiple repeat tests may have been due to specimen mislabeling or contamination during performance of the procedure rather than to the LCx Chlamydia assay. Both assays suffered from a lack of reproducibility on repeat testing with a small proportion of specimens, probably due to the presence of low levels of DNA, the presence of variable amounts of amplification inhibitors, and the loss of DNA during extraction.

Key concepts: Chlamydia trachomatis, Ligase chain reaction, Chlamydia, Chlamydiales, Nucleic Acid Amplification Tests, Urine, Chlamydiaceae, Reproducibility