Comparison of Automated Differential Blood Cell Counts From Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in Normal and Pathologic Samples

Lisa Meintker, J. Ringwald, Manfred Rauh, Stefan W. Krause

Open full text 136 citations

Abstract

Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in a total of 202 samples from hematology patients and normal controls. Manual differential count was used as reference. The analyzers exhibited very good correlation for CBC parameters. Neutrophils and eosinophils also showed very good correlations, whereas lymphocytes and monocytes correlated fairly. The Advia 120 displayed notably lower measurements for both parameters, which is attributable to classification of some events as large unstained cells. Basophil counts were unreliable with all analyzers. Flagging for blasts and immature granulocytes showed moderate sensitivity and specificity. Operators must not rely on blast flagging alone to detect leukemic samples with any analyzer.

Open-access reader

About this research paper

What this paper is about

Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in a total of 202 samples from hematology patients and normal controls. Manual differential count was used as reference. The analyzers exhibited very good correlation for CBC parameters. Neutrophils and eosinophils also showed very good correlations, whereas lymphocytes and monocytes correlated fairly. The Advia 120 displayed notably lower measurements for both parameters, which is attributable to classification of some events as large unstained cells. Basophil counts were unreliable with all analyzers. Flagging for blasts and immature granulocytes showed moderate sensitivity and specificity. Operators must not rely on blast flagging alone to detect leukemic samples with any analyzer.

Why it matters

OpenAlex reports 136 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Reliable automated blood cell characterization and quantification remain challenging in pathologic samples, whereas slide reviews due to unnecessary flagging should be avoided. We compared 4 modern hematology analyzers-Abbott Sapphire, Siemens Advia 120, Sysmex XE-2100, and Beckman Coulter DxH 800-regarding complete blood cell count (CBC), leukocyte differential count, and flagging efficacy in a total of 202 samples from hematology patients and normal controls. Manual differential count was used as reference. The analyzers exhibited very good correlation for CBC parameters. Neutrophils and eosinophils also showed very good correlations, whereas lymphocytes and monocytes correlated fairly. The Advia 120 displayed notably lower measurements for both parameters, which is attributable to classification of some events as large unstained cells. Basophil counts were unreliable with all analyzers. Flagging for blasts and immature granulocytes showed moderate sensitivity and specificity. Operators must not rely on blast flagging alone to detect leukemic samples with any analyzer.

Key concepts: Hematology analyzer, Flagging, Eosinopenia, Medicine, Nuclear medicine, Basophil, Nucleated Red Blood Cell, Pathology