2014•International journal of cardiac imagingOpen access

Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage

Sahar Elahi, Derek S. Ho, Marc D. Feldman, Jouke Dijkstra, Thomas E. Milner

Open full text 1 citations

Abstract

Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250-400 μm) and thin (30-70 μm) phantom neointima. High resolution Micro-CT images of the stent struts were recorded as a gold standard. IVOCT images of the phantom vessels were acquired with various luminal blood scattering strengths and measured neointimal thicknesses from IVOCT and Micro-CT images were compared. In transparent lumen, comparison of IVOCT and Micro-CT neointima thickness measurements found no significant difference (p > 0.05) in the thick neointima phantom but a significant difference (p < 0.05) in the thin neointima phantom. For both thick and thin neointima, IVOCT neointimal thickness measurements varied from Micro-CT values by as much as ±35%. Increased luminal scattering due to presence of blood at concentrations <5% did not interfere with measurement of thin neointimas and was validated by ANOVA analysis (p = 0.95). IV-OCT measurement of strut feature size with an overlying thin neointima match true values determined with Micro-CT (p = 0.82). Presence of a thick neointima resulted in lateral elongation or merry-go-rounding of stent strut features in IVOCT images. Phantom IVOCT images suggest that thick neointimal layers can result in more than 40 % lateral elongation of stent strut features. Simulated IVOCT images of metallic stent struts with varying neointimal thickness suggest that neointimal light scattering can introduce the merry-go-round effect.

Open-access reader

About this research paper

What this paper is about

Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250-400 μm) and thin (30-70 μm) phantom neointima. High resolution Micro-CT images of the stent struts were recorded as a gold standard. IVOCT images of the phantom vessels were acquired with various luminal blood scattering strengths and measured neointimal thicknesses from IVOCT and Micro-CT images were compared. In transparent lumen, comparison of IVOCT and Micro-CT neointima thickness measurements found no significant difference (p > 0.05) in the thick neointima phantom but a significant difference (p < 0.05) in the thin neointima phantom. For both thick and thin neointima, IVOCT neointimal thickness measurements varied from Micro-CT values by as much as ±35%. Increased luminal scattering due to presence of blood at concentrations <5% did not interfere with measurement of thin neointimas and was validated by ANOVA analysis (p = 0.95). IV-OCT measurement of strut feature size with an overlying thin neointima match true values determined with Micro-CT (p = 0.82). Presence of a thick neointima resulted in lateral elongation or merry-go-rounding of stent strut features in IVOCT images. Phantom IVOCT images suggest that thick neointimal layers can result in more than 40 % lateral elongation of stent strut features. Simulated IVOCT images of metallic stent struts with varying neointimal thickness suggest that neointimal light scattering can introduce the merry-go-round effect.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250-400 μm) and thin (30-70 μm) phantom neointima. High resolution Micro-CT images of the stent struts were recorded as a gold standard. IVOCT images of the phantom vessels were acquired with various luminal blood scattering strengths and measured neointimal thicknesses from IVOCT and Micro-CT images were compared. In transparent lumen, comparison of IVOCT and Micro-CT neointima thickness measurements found no significant difference (p > 0.05) in the thick neointima phantom but a significant difference (p < 0.05) in the thin neointima phantom. For both thick and thin neointima, IVOCT neointimal thickness measurements varied from Micro-CT values by as much as ±35%. Increased luminal scattering due to presence of blood at concentrations <5% did not interfere with measurement of thin neointimas and was validated by ANOVA analysis (p = 0.95). IV-OCT measurement of strut feature size with an overlying thin neointima match true values determined with Micro-CT (p = 0.82). Presence of a thick neointima resulted in lateral elongation or merry-go-rounding of stent strut features in IVOCT images. Phantom IVOCT images suggest that thick neointimal layers can result in more than 40 % lateral elongation of stent strut features. Simulated IVOCT images of metallic stent struts with varying neointimal thickness suggest that neointimal light scattering can introduce the merry-go-round effect.

Key concepts: Neointima, Imaging phantom, Stent, Biomedical engineering, Lumen (anatomy), Medicine, Coronary stent, Materials science

Related papers

Back to paper searchBrowse research topicsOriginal source
Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage — Research Paper | ScholarLens