Assessment of sequential same arm agreement of blood pressure measurements by a CVProfilor™ DO–2020 versus a Baumanometer® mercury sphygmomanometer
L. Michael Prisant, Lawrence M. Resnick, Steven M. Hollenberg
Abstract
L. Michael Prisant, Lawrence M. Resnick, Steven M. Hollenberg
Abstract
OBJECTIVE: The aim of this study was to assess the accuracy of sequential same arm blood pressure measurement by the mercury sphygmomanometer with the oscillometric blood pressure measurements from a device that also determines arterial elasticity. METHODS: A prospective, multicentre, clinical study evaluated sequential same arm blood pressure measurements, using a mercury sphygmomanometer (Baumanometer, W. A. Baum Co., Inc., Copiague, New York, USA) and an oscillometric non-invasive device that calculates arterial elasticity (CVProfilor DO-2020 Cardiovascular Profiling System, Hypertension Diagnostics, Inc., Eagan, Minnesota, USA). Blood pressure was measured supine in triplicate, 3 min apart in a randomized sequence after a period of rest. RESULTS: The study population of 230 normotensive and hypertensive subjects included 57% females, 51% Caucasians, and 33% African Americans. The mean difference between test methods of systolic blood pressure, diastolic blood pressure, and heart rate was -3.2 +/- 6.9 mmHg, +0.8 +/- 5.9 mmHg, and +1.0 +/- 5.7 beats/minute. For systolic and diastolic blood pressure, 60.9 and 70.4% of sequential measurements by each method were within +/- 5 mmHg. Few or no points fell beyond the mean +/- 2 standard deviations lines for each cuff bladder size. CONCLUSION: Sequential same arm measurements of the CVProfilor DO-2020 Cardiovascular Profiling System measures blood pressure by an oscillometric method (dynamic linear deflation) with reasonable agreement with a mercury sphygmomanometer.
OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
OBJECTIVE: The aim of this study was to assess the accuracy of sequential same arm blood pressure measurement by the mercury sphygmomanometer with the oscillometric blood pressure measurements from a device that also determines arterial elasticity. METHODS: A prospective, multicentre, clinical study evaluated sequential same arm blood pressure measurements, using a mercury sphygmomanometer (Baumanometer, W. A. Baum Co., Inc., Copiague, New York, USA) and an oscillometric non-invasive device that calculates arterial elasticity (CVProfilor DO-2020 Cardiovascular Profiling System, Hypertension Diagnostics, Inc., Eagan, Minnesota, USA). Blood pressure was measured supine in triplicate, 3 min apart in a randomized sequence after a period of rest. RESULTS: The study population of 230 normotensive and hypertensive subjects included 57% females, 51% Caucasians, and 33% African Americans. The mean difference between test methods of systolic blood pressure, diastolic blood pressure, and heart rate was -3.2 +/- 6.9 mmHg, +0.8 +/- 5.9 mmHg, and +1.0 +/- 5.7 beats/minute. For systolic and diastolic blood pressure, 60.9 and 70.4% of sequential measurements by each method were within +/- 5 mmHg. Few or no points fell beyond the mean +/- 2 standard deviations lines for each cuff bladder size. CONCLUSION: Sequential same arm measurements of the CVProfilor DO-2020 Cardiovascular Profiling System measures blood pressure by an oscillometric method (dynamic linear deflation) with reasonable agreement with a mercury sphygmomanometer.
Key concepts: Sphygmomanometer, Medicine, Blood pressure, Supine position, Cuff, Cardiology, Diastole, Internal medicine