1999Analytical LettersRequires access

Ion-Selective Electrode for the Determination of Furaltadone

Sabry Khalil, S. Abd El-Aliem

Open publisher page 1 citations

Abstract

A Furaltadone (FD) ion-selective PVC membrane electrode based on the ion-pair complex of FD with sodium tetraphenylborate was prepared with dioctyl phthalate as a plasticiser. The electrode exhibits a linear response with an approximate Nernstian slope (50 mV decade−1 at 20 0C) with the concentration range 2.18 x 10−5–2.18 x 10−4 FD. The effects of ionic strength and pH of the test solution on the electrode performance were studied. The electrode exhibited very good selectivity for FD with respect to a large number of inorganic and organic cations of biological importance. The standard addition method and potentiometric titration were used to determine the FD in pure solution and in pharmaceutical preparations with satisfactory results.

About this research paper

What this paper is about

A Furaltadone (FD) ion-selective PVC membrane electrode based on the ion-pair complex of FD with sodium tetraphenylborate was prepared with dioctyl phthalate as a plasticiser. The electrode exhibits a linear response with an approximate Nernstian slope (50 mV decade−1 at 20 0C) with the concentration range 2.18 x 10−5–2.18 x 10−4 FD. The effects of ionic strength and pH of the test solution on the electrode performance were studied. The electrode exhibited very good selectivity for FD with respect to a large number of inorganic and organic cations of biological importance. The standard addition method and potentiometric titration were used to determine the FD in pure solution and in pharmaceutical preparations with satisfactory results.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A Furaltadone (FD) ion-selective PVC membrane electrode based on the ion-pair complex of FD with sodium tetraphenylborate was prepared with dioctyl phthalate as a plasticiser. The electrode exhibits a linear response with an approximate Nernstian slope (50 mV decade−1 at 20 0C) with the concentration range 2.18 x 10−5–2.18 x 10−4 FD. The effects of ionic strength and pH of the test solution on the electrode performance were studied. The electrode exhibited very good selectivity for FD with respect to a large number of inorganic and organic cations of biological importance. The standard addition method and potentiometric titration were used to determine the FD in pure solution and in pharmaceutical preparations with satisfactory results.

Key concepts: Chemistry, Potentiometric titration, Sodium tetraphenylborate, Tetraphenylborate, Ion selective electrode, Electrode, Plasticizer, Inorganic chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
Ion-Selective Electrode for the Determination of Furaltadone — Research Paper | ScholarLens