Comparison of four immunoassay analyzers for relationship between thyroid stimulating hormone (TSH) and free thyroxine (FT4)

Muhittin A. Serdar, Emre İspir, Taner Özgürtaş, Özlem Gülbahar, Zahid Ciraci, Hatîce Paşaoğlu, İsmail Kurt

Open full text 4 citations

Abstract

Objective: There is an inverse log/linear relationship between TSH and FT4 due to the negative feedback of these hormones on the pituitary. The objective of our study was to compare this relationship of TSH and FT4 between four different immunoassay analyzers. Methods: In our study, four data sets obtained from the database of four different hospital laboratories each using only a single method of measurement was used to evaluate this relationship between TSH and FT4.These data sets with their assay methods include: 21.102 test results measured by Modular E170 Analyzer (Roche Diagnostics, Germany), 20.241 test results measured by Access DxI 800 Unicel (Beckman Coulter, USA), 22.444 test results measured by Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) and 20.200 test results measured by ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY). Inverse logarithmic relationship were determined from each data sets and compared between analyzers. Results: The correlation coefficients were -0.439 [95% CI, (-0.450)- (-0.428)], -0.488 [95% CI, (-0.498)- (-0.478)], -0.353 [95% CI, (-0.364)- (-0.342)], -0.430 [95% CI, (-0.441)- (-0.419)] for DxI 800 Unicel, Modular E 170, Architect i2000sr and ADVIA Centaur XP, respectively. In our study, all immunoassay analyzers showed poor correlation in a concentration range of TSH between 1.0 to 10.0 uIU/mL and the inverse log/linear relationship was not observed. Inverse relationship between TSH and FT4 is the result of negative thyroid pituitary hypothalamic feedback mechanism and this relationship is important especially in the diagnosis and treatment of hypo and hyperthyroidism. In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4. Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation.

Open-access reader

About this research paper

What this paper is about

Objective: There is an inverse log/linear relationship between TSH and FT4 due to the negative feedback of these hormones on the pituitary. The objective of our study was to compare this relationship of TSH and FT4 between four different immunoassay analyzers. Methods: In our study, four data sets obtained from the database of four different hospital laboratories each using only a single method of measurement was used to evaluate this relationship between TSH and FT4.These data sets with their assay methods include: 21.102 test results measured by Modular E170 Analyzer (Roche Diagnostics, Germany), 20.241 test results measured by Access DxI 800 Unicel (Beckman Coulter, USA), 22.444 test results measured by Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) and 20.200 test results measured by ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY). Inverse logarithmic relationship were determined from each data sets and compared between analyzers. Results: The correlation coefficients were -0.439 [95% CI, (-0.450)- (-0.428)], -0.488 [95% CI, (-0.498)- (-0.478)], -0.353 [95% CI, (-0.364)- (-0.342)], -0.430 [95% CI, (-0.441)- (-0.419)] for DxI 800 Unicel, Modular E 170, Architect i2000sr and ADVIA Centaur XP, respectively. In our study, all immunoassay analyzers showed poor correlation in a concentration range of TSH between 1.0 to 10.0 uIU/mL and the inverse log/linear relationship was not observed. Inverse relationship between TSH and FT4 is the result of negative thyroid pituitary hypothalamic feedback mechanism and this relationship is important especially in the diagnosis and treatment of hypo and hyperthyroidism. In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4. Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation.

Why it matters

OpenAlex reports 4 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective: There is an inverse log/linear relationship between TSH and FT4 due to the negative feedback of these hormones on the pituitary. The objective of our study was to compare this relationship of TSH and FT4 between four different immunoassay analyzers. Methods: In our study, four data sets obtained from the database of four different hospital laboratories each using only a single method of measurement was used to evaluate this relationship between TSH and FT4.These data sets with their assay methods include: 21.102 test results measured by Modular E170 Analyzer (Roche Diagnostics, Germany), 20.241 test results measured by Access DxI 800 Unicel (Beckman Coulter, USA), 22.444 test results measured by Architect i2000sr (Abbott Laboratories, Abbott Park, Illinois, U.S.A) and 20.200 test results measured by ADVIA Centaur XP (Siemens Diagnostics, Tarrytown, NY). Inverse logarithmic relationship were determined from each data sets and compared between analyzers. Results: The correlation coefficients were -0.439 [95% CI, (-0.450)- (-0.428)], -0.488 [95% CI, (-0.498)- (-0.478)], -0.353 [95% CI, (-0.364)- (-0.342)], -0.430 [95% CI, (-0.441)- (-0.419)] for DxI 800 Unicel, Modular E 170, Architect i2000sr and ADVIA Centaur XP, respectively. In our study, all immunoassay analyzers showed poor correlation in a concentration range of TSH between 1.0 to 10.0 uIU/mL and the inverse log/linear relationship was not observed. Inverse relationship between TSH and FT4 is the result of negative thyroid pituitary hypothalamic feedback mechanism and this relationship is important especially in the diagnosis and treatment of hypo and hyperthyroidism. In data sets obtained from four different immunoassay analyzers, poor and statistically different correlation was observed between analyzers at TSH values ranged from 1.0 to 10.0 uIU/mL compared with FT4. Conclusion: These variations between analyzers may affect the clinical decisions especially in the evaluation of subclinical hypothyroidism, clinicians and laboratory specialists should be aware of these situation.

Key concepts: Immunoassay, Thyroid-stimulating hormone, Free thyroxine, Internal medicine, Endocrinology, Hormone, Linear relationship, Thyroid hormones

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of four immunoassay analyzers for relationship between thyroid stimulating hormone (TSH) and free thyroxine (FT4) — Research Paper | ScholarLens