Comparison of two second‐generation anti‐hepatitis C virus ELISA on 21431 US blood donor samples
Steven Anderson, Tanya R. Hathaway, I. K. Kuramoto, P. V. Holland, R. Gilchec, Tamar Koch, Sally Hojvat
Abstract
Steven Anderson, Tanya R. Hathaway, I. K. Kuramoto, P. V. Holland, R. Gilchec, Tamar Koch, Sally Hojvat
Abstract
We have compared two different second-generation (2.0) enzyme-linked immunosorbent assays (ELISA) for the presence of antibodies to hepatitis C virus (anti-HCV) in blood from volunteer, unpaid donors. At two separate blood centres, a total of 21,431 donor samples were tested with Abbott Anti-HCV 2.0 ELISA and Ortho Anti-HCV 2.0 ELISA. Samples found to be repeatedly reactive were tested by supplemental/investigational assays. MATRIX HCV (Abbott) and anti-HCV RIBA II (Ortho/Chiron), to 'confirm' the presence of anti-HCV. Discordant ELISA samples were additionally tested by the polymerase chain reaction (PCR) for the presence of HCV RNA. The Abbott anti-HCV assay had a repeatedly reactive rate of 0.59% (127/21,431) and the Ortho anti-HCV assay 0.51% (110/21,431). Overall agreement between assays was 99.76%, 72/127 (56.7%) of Abbott repeatedly reactive samples confirmed on MATRIX and 61/127 (48.0%) on RIBAII; 70/110 (63.6%) of Ortho repeatedly reactivate samples confirmed on MATRIX and 61/110 (55.5%) on RIBA II. Discordant ELISA samples tested by PCR yielded negative results. Hence the two ELISA had equal sensitivity, as defined by detection of true positive samples; the slightly lower specificity of the Abbott Anti-HCV 2.0 ELISA may be owing to culling of donors with a false positive test by Ortho's Anti-HCV 1.0 and 2.0 ELISA tests (the routine tests in place at each blood centre). A sample found to be repeatedly reactive by two different ELISA tests for anti-HCV is likely to be a true positive and may not require further 'confirmatory' testing.
OpenAlex reports 11 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
We have compared two different second-generation (2.0) enzyme-linked immunosorbent assays (ELISA) for the presence of antibodies to hepatitis C virus (anti-HCV) in blood from volunteer, unpaid donors. At two separate blood centres, a total of 21,431 donor samples were tested with Abbott Anti-HCV 2.0 ELISA and Ortho Anti-HCV 2.0 ELISA. Samples found to be repeatedly reactive were tested by supplemental/investigational assays. MATRIX HCV (Abbott) and anti-HCV RIBA II (Ortho/Chiron), to 'confirm' the presence of anti-HCV. Discordant ELISA samples were additionally tested by the polymerase chain reaction (PCR) for the presence of HCV RNA. The Abbott anti-HCV assay had a repeatedly reactive rate of 0.59% (127/21,431) and the Ortho anti-HCV assay 0.51% (110/21,431). Overall agreement between assays was 99.76%, 72/127 (56.7%) of Abbott repeatedly reactive samples confirmed on MATRIX and 61/127 (48.0%) on RIBAII; 70/110 (63.6%) of Ortho repeatedly reactivate samples confirmed on MATRIX and 61/110 (55.5%) on RIBA II. Discordant ELISA samples tested by PCR yielded negative results. Hence the two ELISA had equal sensitivity, as defined by detection of true positive samples; the slightly lower specificity of the Abbott Anti-HCV 2.0 ELISA may be owing to culling of donors with a false positive test by Ortho's Anti-HCV 1.0 and 2.0 ELISA tests (the routine tests in place at each blood centre). A sample found to be repeatedly reactive by two different ELISA tests for anti-HCV is likely to be a true positive and may not require further 'confirmatory' testing.
Key concepts: Medicine, Hepatitis C virus, Virology, Blood donor, Antibody, Serology, Hepatitis C, Virus