Isotope ratio and isotope dilution measurements using axial inductively coupled plasma time of flight mass spectrometry

Håkan Emteborg, Xiaodan Tian, Markus Ostermann, Michael Berglund, Freddy C. Adams

Open publisher page 46 citations

Abstract

Performance of axial ICP-TOF-MS in terms of precision and mass bias for isotope ratio measurements and accuracy of isotope dilution was assessed by measurements of isotopic reference materials for Li, Mg, Rb, Pt and Pb. Because of simultaneous ion extraction from the plasma, remaining flicker noise is minimized resulting in improved precision for isotope ratios compared to sequential mass spectrometers. Experimental relative standard deviations of <0.05% for isotope ratios were obtained at high signal levels measured in the analog detection mode. Effects on isotope ratios derived from changes of instrumental parameters, such as detector voltage and transverse rejection ion pulse settings, were evaluated. Isotope ratios for different concentrations changed when operating at too low detector voltages. An appropriate detector voltage setting is important but optimum voltages differ depending on the age and previous history of the detector. Results obtained for ID measurements of magnesium and rubidium amount content compared well with results from quadrupole ICP-MS and thermal ionization mass spectrometry. Mass bias per mass unit is high in the low mass range (13% forLi/Li) but around 0.2% in the high mass range. The long-term stability of ratios, measured discontinuously over several hours, is moderate; therefore measurements of an isotopic reference material at regular intervals is necessary to correct for small variations in mass bias over time.

About this research paper

What this paper is about

Performance of axial ICP-TOF-MS in terms of precision and mass bias for isotope ratio measurements and accuracy of isotope dilution was assessed by measurements of isotopic reference materials for Li, Mg, Rb, Pt and Pb. Because of simultaneous ion extraction from the plasma, remaining flicker noise is minimized resulting in improved precision for isotope ratios compared to sequential mass spectrometers. Experimental relative standard deviations of <0.05% for isotope ratios were obtained at high signal levels measured in the analog detection mode. Effects on isotope ratios derived from changes of instrumental parameters, such as detector voltage and transverse rejection ion pulse settings, were evaluated. Isotope ratios for different concentrations changed when operating at too low detector voltages. An appropriate detector voltage setting is important but optimum voltages differ depending on the age and previous history of the detector. Results obtained for ID measurements of magnesium and rubidium amount content compared well with results from quadrupole ICP-MS and thermal ionization mass spectrometry. Mass bias per mass unit is high in the low mass range (13% forLi/Li) but around 0.2% in the high mass range. The long-term stability of ratios, measured discontinuously over several hours, is moderate; therefore measurements of an isotopic reference material at regular intervals is necessary to correct for small variations in mass bias over time.

Why it matters

OpenAlex reports 46 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Performance of axial ICP-TOF-MS in terms of precision and mass bias for isotope ratio measurements and accuracy of isotope dilution was assessed by measurements of isotopic reference materials for Li, Mg, Rb, Pt and Pb. Because of simultaneous ion extraction from the plasma, remaining flicker noise is minimized resulting in improved precision for isotope ratios compared to sequential mass spectrometers. Experimental relative standard deviations of <0.05% for isotope ratios were obtained at high signal levels measured in the analog detection mode. Effects on isotope ratios derived from changes of instrumental parameters, such as detector voltage and transverse rejection ion pulse settings, were evaluated. Isotope ratios for different concentrations changed when operating at too low detector voltages. An appropriate detector voltage setting is important but optimum voltages differ depending on the age and previous history of the detector. Results obtained for ID measurements of magnesium and rubidium amount content compared well with results from quadrupole ICP-MS and thermal ionization mass spectrometry. Mass bias per mass unit is high in the low mass range (13% forLi/Li) but around 0.2% in the high mass range. The long-term stability of ratios, measured discontinuously over several hours, is moderate; therefore measurements of an isotopic reference material at regular intervals is necessary to correct for small variations in mass bias over time.

Key concepts: Chemistry, Mass spectrometry, Analytical Chemistry (journal), Isotope, Isotope dilution, Thermal ionization mass spectrometry, Inductively coupled plasma mass spectrometry, Quadrupole mass analyzer

Related papers

Back to paper searchBrowse research topicsOriginal source
Isotope ratio and isotope dilution measurements using axial inductively coupled plasma time of flight mass spectrometry — Research Paper | ScholarLens