A novel balloon angioplasty catheter impregnated with beta-particle emitting radioisotopes for vascular brachytherapy to prevent restenosis. First in vivo results

Christoph Hehrlein

Open publisher page 15 citations

Abstract

BACKGROUND: According to early clinical trials, vascular brachytherapy performed prior to or shortly after angioplasty is very effective in reducing restenosis rates. The purpose of this study was to investigate the effects of a novel radioactive catheter that allows simultaneous balloon angioplasty and beta-particle irradiation in the prevention of restenosis. MATERIAL AND METHODS: The balloon surface of an angioplasty catheter was impregnated with the radioisotope(32)P. Dosimetry calculations using a Monte Carlo method were performed at a radial distance of 0.2 mm from the balloon surface. Rabbit iliac arteries were dilated and simultaneously irradiated with a dose of 20 Gy delivered to the adventitia. Control arteries were only dilated and not irradiated. Neointimal areas, cell numbers and the perimeter of the arteries were measured by histomorphometry after 6 weeks. RESULTS: Neointima formation was reduced after balloon dilatation and simultaneous beta-particle irradiation using the(32)P impregnated angioplasty catheter as compared to balloon dilatation alone with a non-impregnated catheter (0.09+/-0.06 vs 0.27+/-0.09 mm(2)neointimal area and 168+/-45 vs 360+/-133 cells/0.05 mm(2)neointima, P<0.001 vs control, respectively). In addition, balloon dilatation with the(32)P impregnated angioplasty catheter increased the vessel perimeter as compared to balloon dilatation with a non-impregnated catheter (4. 7+/-0.2 vs 3.9+/-0.3 mm, P<0.001 vs control). CONCLUSIONS: Simultaneous balloon dilatation and vascular brachytherapy with a novel(32)P impregnated angioplasty catheter markedly reduces restenosis in vivo by preventing neointimal hyperplasia and constrictive vascular remodelling.

About this research paper

What this paper is about

BACKGROUND: According to early clinical trials, vascular brachytherapy performed prior to or shortly after angioplasty is very effective in reducing restenosis rates. The purpose of this study was to investigate the effects of a novel radioactive catheter that allows simultaneous balloon angioplasty and beta-particle irradiation in the prevention of restenosis. MATERIAL AND METHODS: The balloon surface of an angioplasty catheter was impregnated with the radioisotope(32)P. Dosimetry calculations using a Monte Carlo method were performed at a radial distance of 0.2 mm from the balloon surface. Rabbit iliac arteries were dilated and simultaneously irradiated with a dose of 20 Gy delivered to the adventitia. Control arteries were only dilated and not irradiated. Neointimal areas, cell numbers and the perimeter of the arteries were measured by histomorphometry after 6 weeks. RESULTS: Neointima formation was reduced after balloon dilatation and simultaneous beta-particle irradiation using the(32)P impregnated angioplasty catheter as compared to balloon dilatation alone with a non-impregnated catheter (0.09+/-0.06 vs 0.27+/-0.09 mm(2)neointimal area and 168+/-45 vs 360+/-133 cells/0.05 mm(2)neointima, P<0.001 vs control, respectively). In addition, balloon dilatation with the(32)P impregnated angioplasty catheter increased the vessel perimeter as compared to balloon dilatation with a non-impregnated catheter (4. 7+/-0.2 vs 3.9+/-0.3 mm, P<0.001 vs control). CONCLUSIONS: Simultaneous balloon dilatation and vascular brachytherapy with a novel(32)P impregnated angioplasty catheter markedly reduces restenosis in vivo by preventing neointimal hyperplasia and constrictive vascular remodelling.

Why it matters

OpenAlex reports 15 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: According to early clinical trials, vascular brachytherapy performed prior to or shortly after angioplasty is very effective in reducing restenosis rates. The purpose of this study was to investigate the effects of a novel radioactive catheter that allows simultaneous balloon angioplasty and beta-particle irradiation in the prevention of restenosis. MATERIAL AND METHODS: The balloon surface of an angioplasty catheter was impregnated with the radioisotope(32)P. Dosimetry calculations using a Monte Carlo method were performed at a radial distance of 0.2 mm from the balloon surface. Rabbit iliac arteries were dilated and simultaneously irradiated with a dose of 20 Gy delivered to the adventitia. Control arteries were only dilated and not irradiated. Neointimal areas, cell numbers and the perimeter of the arteries were measured by histomorphometry after 6 weeks. RESULTS: Neointima formation was reduced after balloon dilatation and simultaneous beta-particle irradiation using the(32)P impregnated angioplasty catheter as compared to balloon dilatation alone with a non-impregnated catheter (0.09+/-0.06 vs 0.27+/-0.09 mm(2)neointimal area and 168+/-45 vs 360+/-133 cells/0.05 mm(2)neointima, P<0.001 vs control, respectively). In addition, balloon dilatation with the(32)P impregnated angioplasty catheter increased the vessel perimeter as compared to balloon dilatation with a non-impregnated catheter (4. 7+/-0.2 vs 3.9+/-0.3 mm, P<0.001 vs control). CONCLUSIONS: Simultaneous balloon dilatation and vascular brachytherapy with a novel(32)P impregnated angioplasty catheter markedly reduces restenosis in vivo by preventing neointimal hyperplasia and constrictive vascular remodelling.

Key concepts: Neointima, Restenosis, Medicine, Angioplasty, Balloon, Brachytherapy, Neointimal hyperplasia, Balloon catheter