Determination of lead and cadmium in tea by graphite furnace atomic absorption
MU Hai-yan, Zhonglan Chen
Abstract
Open-access reader
MU Hai-yan, Zhonglan Chen
Abstract
Open-access reader
Abstract Lead and cadmium were determined in the presence of matrix modifiers. Under the optimal measure conditions, lead and cadmium in tea showed excellent linearities with correlation coefficients of 0.9996 and 0.9993 after digestion of tea with HNO 3 /HCl (3:1, v/v). The recoveries were 94–102% and 91–103% and RSD values were 8.86% and 13.1%, respectively.
OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Abstract Lead and cadmium were determined in the presence of matrix modifiers. Under the optimal measure conditions, lead and cadmium in tea showed excellent linearities with correlation coefficients of 0.9996 and 0.9993 after digestion of tea with HNO 3 /HCl (3:1, v/v). The recoveries were 94–102% and 91–103% and RSD values were 8.86% and 13.1%, respectively.
Key concepts: Cadmium, Graphite furnace atomic absorption, Lead (geology), Chemistry, Atomic absorption spectroscopy, Matrix (chemical analysis), Graphite, Absorption (acoustics)