Dual-Energy X-Ray Absorptiometry for Measurement of Phalangeal Bone Mineral Density on a Slot-Scanning Digital Radiography System

Ronald Dendere, J.H. Potgieter, Stefan H. Steiner, S. P. Whiley, Tania S. Douglas

Open publisher page 14 citations

Abstract

OBJECTIVE: In this paper, we assess the feasibility of using two detectors in a slot-scanning digital radiography system to acquire images for measuring bone mineral density (BMD) of the middle phalanx of the middle finger using dual-energy X-ray absorptiometry (DXA). METHODS: Simulations were used to evaluate the spectral separation of the low- and high-energy spectra and detective quantum efficiency was used for assessing image quality. Scan parameters were chosen to optimize spectral separation, image quality, and radiation dose. We introduce the measurement of volumetric BMD (vBMD) using basis material decomposition. We assess the accuracy of our methods by comparing measurements taken using bone images against reference data derived from subsequent incineration of the bones. In vivo scans were conducted to evaluate the system precision (repeatability) and agreement with a clinical densitometer. RESULTS: Average errors for bone mineral content (BMC), areal BMD (aBMD), and vBMD were 4.85%, 5.49%, and 12.77%, respectively. Our system had good agreement with a clinical densitometer based on concordance correlation coefficient values of 0.92 and 0.98 for aBMD and BMC, respectively. Precision studies yielded coefficient of variation (CV) values of 1.35% for aBMD, 1.48% for BMC, and 1.80% for vBMD. The CV values of all measurements were within 2%, indicating that the methods have clinically acceptable precision. CONCLUSION: We conclude that our techniques yield bone measurements with high accuracy, clinically acceptable precision, and good agreement with a clinical densitometer. SIGNIFICANCE: We have shown the clinical potential of phalangeal DXA measurements of aBMD and vBMD on a slot-scanning digital radiography system.

About this research paper

What this paper is about

OBJECTIVE: In this paper, we assess the feasibility of using two detectors in a slot-scanning digital radiography system to acquire images for measuring bone mineral density (BMD) of the middle phalanx of the middle finger using dual-energy X-ray absorptiometry (DXA). METHODS: Simulations were used to evaluate the spectral separation of the low- and high-energy spectra and detective quantum efficiency was used for assessing image quality. Scan parameters were chosen to optimize spectral separation, image quality, and radiation dose. We introduce the measurement of volumetric BMD (vBMD) using basis material decomposition. We assess the accuracy of our methods by comparing measurements taken using bone images against reference data derived from subsequent incineration of the bones. In vivo scans were conducted to evaluate the system precision (repeatability) and agreement with a clinical densitometer. RESULTS: Average errors for bone mineral content (BMC), areal BMD (aBMD), and vBMD were 4.85%, 5.49%, and 12.77%, respectively. Our system had good agreement with a clinical densitometer based on concordance correlation coefficient values of 0.92 and 0.98 for aBMD and BMC, respectively. Precision studies yielded coefficient of variation (CV) values of 1.35% for aBMD, 1.48% for BMC, and 1.80% for vBMD. The CV values of all measurements were within 2%, indicating that the methods have clinically acceptable precision. CONCLUSION: We conclude that our techniques yield bone measurements with high accuracy, clinically acceptable precision, and good agreement with a clinical densitometer. SIGNIFICANCE: We have shown the clinical potential of phalangeal DXA measurements of aBMD and vBMD on a slot-scanning digital radiography system.

Why it matters

OpenAlex reports 14 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: In this paper, we assess the feasibility of using two detectors in a slot-scanning digital radiography system to acquire images for measuring bone mineral density (BMD) of the middle phalanx of the middle finger using dual-energy X-ray absorptiometry (DXA). METHODS: Simulations were used to evaluate the spectral separation of the low- and high-energy spectra and detective quantum efficiency was used for assessing image quality. Scan parameters were chosen to optimize spectral separation, image quality, and radiation dose. We introduce the measurement of volumetric BMD (vBMD) using basis material decomposition. We assess the accuracy of our methods by comparing measurements taken using bone images against reference data derived from subsequent incineration of the bones. In vivo scans were conducted to evaluate the system precision (repeatability) and agreement with a clinical densitometer. RESULTS: Average errors for bone mineral content (BMC), areal BMD (aBMD), and vBMD were 4.85%, 5.49%, and 12.77%, respectively. Our system had good agreement with a clinical densitometer based on concordance correlation coefficient values of 0.92 and 0.98 for aBMD and BMC, respectively. Precision studies yielded coefficient of variation (CV) values of 1.35% for aBMD, 1.48% for BMC, and 1.80% for vBMD. The CV values of all measurements were within 2%, indicating that the methods have clinically acceptable precision. CONCLUSION: We conclude that our techniques yield bone measurements with high accuracy, clinically acceptable precision, and good agreement with a clinical densitometer. SIGNIFICANCE: We have shown the clinical potential of phalangeal DXA measurements of aBMD and vBMD on a slot-scanning digital radiography system.

Key concepts: Densitometer, Bone mineral, Nuclear medicine, Digital radiography, Accuracy and precision, Repeatability, Densitometry, Radiography