A New Ultrasonic Transducer to Decrease the Multi-Reflections Which Cause Artifacts on B-Mode Imaging
F. Namiki, A. Iida, T. Shimura
Abstract
F. Namiki, A. Iida, T. Shimura
Abstract
We propose a design that decreases the acoustic energy reflected by an ultrasonic t ransducer f or medical use. Multi-reflections must be reduced to achieve good images without artifacts. This can be done by decreasing the transducer's r eflection coefficient. The reflection coefficient is important, and must be taken into a ccount t ogether with efficiency and bandwidth when considering design. Input acoustic impedance is assumed to be that seen from the transducer's surface to the backing material through the matching layers a nd piezoelectric ceramic (PZT). Input acoustic impedance at the center frequency is designed to equal t he acoustic impedance of the load. We analized the reflection c oefficients of several transducers by computer simulation. We found a structure that provides a small reflection coefficient, good efficiency, and a wide bandwidth. The reflection coefficient of the thin-disk transducer we fabricated is 10 dB lower than that of a conventional transducer.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
We propose a design that decreases the acoustic energy reflected by an ultrasonic t ransducer f or medical use. Multi-reflections must be reduced to achieve good images without artifacts. This can be done by decreasing the transducer's r eflection coefficient. The reflection coefficient is important, and must be taken into a ccount t ogether with efficiency and bandwidth when considering design. Input acoustic impedance is assumed to be that seen from the transducer's surface to the backing material through the matching layers a nd piezoelectric ceramic (PZT). Input acoustic impedance at the center frequency is designed to equal t he acoustic impedance of the load. We analized the reflection c oefficients of several transducers by computer simulation. We found a structure that provides a small reflection coefficient, good efficiency, and a wide bandwidth. The reflection coefficient of the thin-disk transducer we fabricated is 10 dB lower than that of a conventional transducer.
Key concepts: Transducer, Reflection coefficient, Acoustic impedance, Acoustics, Ultrasonic sensor, Impedance matching, Materials science, Bandwidth (computing)