Assessment of the Trueness and Inter-Laboratory Precision of Routine Uric Acid Assays Using 4 Frozen Pooled Serum Samples Measured by the Japan Society of Clinical Chemistry's HPLC Method

Yaping Jiang, Ou Liu, Guobing Xu

Open full text 3 citations

Abstract

BACKGROUND: Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical systems using a candidate reference method. METHODS: We performed the HPLC method recommended by the Japan Society of Clinical Chemistry (JSCC). Next, we evaluated its analytical performance and validated its trueness. The performance of 4 routine analytical systems (5 instruments per system, n=20) for UA was assessed by using 4 frozen pooled serum samples measured by the HPLC method according to biologically relevant quality goals. RESULTS: Within-run, between-run, inter-day, and total CV of the method were less than 0.3%, 0.4%, 1.8%, and 2.6%, respectively. The UA measurements were consistent with the target values of standard reference material (SRM) 909b, the sixth ring trial for Reference Laboratories (RELA-2008) specimen, and national primary reference materials. The 4 frozen pooled serum samples were homogeneous, stable, and commutable. All routine systems achieved the desirable performance goal (total error <11.9%). CONCLUSIONS: We successfully reproduced the JSCC's HPLC method, which was simple, specific, precise, and accurate. We recommend this method as a reference method for UA measurement in human serum. Four routine analytical systems for UA measurement had acceptable traceability, and their UA results showed good concordance.

Open-access reader

About this research paper

What this paper is about

BACKGROUND: Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical systems using a candidate reference method. METHODS: We performed the HPLC method recommended by the Japan Society of Clinical Chemistry (JSCC). Next, we evaluated its analytical performance and validated its trueness. The performance of 4 routine analytical systems (5 instruments per system, n=20) for UA was assessed by using 4 frozen pooled serum samples measured by the HPLC method according to biologically relevant quality goals. RESULTS: Within-run, between-run, inter-day, and total CV of the method were less than 0.3%, 0.4%, 1.8%, and 2.6%, respectively. The UA measurements were consistent with the target values of standard reference material (SRM) 909b, the sixth ring trial for Reference Laboratories (RELA-2008) specimen, and national primary reference materials. The 4 frozen pooled serum samples were homogeneous, stable, and commutable. All routine systems achieved the desirable performance goal (total error <11.9%). CONCLUSIONS: We successfully reproduced the JSCC's HPLC method, which was simple, specific, precise, and accurate. We recommend this method as a reference method for UA measurement in human serum. Four routine analytical systems for UA measurement had acceptable traceability, and their UA results showed good concordance.

Why it matters

OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: Reference procedures are required for evaluating the accuracy of routine analytical systems for uric acid (UA). External quality assessment (EQA) for UA has only been conducted with quality controls in China, and the results have not been published. This study was designed to investigate both the trueness and inter-laboratory precision of UA measurements among routine analytical systems using a candidate reference method. METHODS: We performed the HPLC method recommended by the Japan Society of Clinical Chemistry (JSCC). Next, we evaluated its analytical performance and validated its trueness. The performance of 4 routine analytical systems (5 instruments per system, n=20) for UA was assessed by using 4 frozen pooled serum samples measured by the HPLC method according to biologically relevant quality goals. RESULTS: Within-run, between-run, inter-day, and total CV of the method were less than 0.3%, 0.4%, 1.8%, and 2.6%, respectively. The UA measurements were consistent with the target values of standard reference material (SRM) 909b, the sixth ring trial for Reference Laboratories (RELA-2008) specimen, and national primary reference materials. The 4 frozen pooled serum samples were homogeneous, stable, and commutable. All routine systems achieved the desirable performance goal (total error <11.9%). CONCLUSIONS: We successfully reproduced the JSCC's HPLC method, which was simple, specific, precise, and accurate. We recommend this method as a reference method for UA measurement in human serum. Four routine analytical systems for UA measurement had acceptable traceability, and their UA results showed good concordance.