1983•ThoraxOpen access

Errors in the measurement of total lung capacity in chronic obstructive lung disease.

Peter D. Paré, Barry Wiggs, Catherine A. Coppin

Open full text 27 citations

Abstract

The standard plethysmographic method of measuring total lung capacity (TLC) has been reported to result in spuriously high estimates in patients with severe airway obstruction. The helium-dilution method is known to underestimate TLC in the same patients. To determine the magnitude of these possible errors we measured TLC by four methods in 20 patients with varying degrees of chronic obstructive lung disease and in 11 normal subjects. TLC was measured by (1) helium dilution (TLCHe); (2) a volume-displacement body plethysmograph, box volume being plotted against mouth pressure (TLCm); (3) the same body plethysmograph with volume plotted against pressure measured with an oesophageal balloon (TLCes); and (4) a radiological technique (TLCxr). In normal subjects there was no difference between TCLm (6.57 +/- 1.20) and TLCes (6.51 +/- 1.24). In the patients with chronic obstructive lung disease TLCm gave results significantly higher than those of any other method. If TLCes is taken as the closest estimate of true TLC, TLCm consistently overestimates and TLCHe underestimates TLC. There was no relationship between the degree of airway obstruction and (TLCm - TLCes) but there was between (TLCes - TLCHe) and severity of airway obstruction. We conclude that using mouth pressure in the plethysmographic measurement of TLC in patients with chronic obstructive lung disease results in consistent but slight overestimation of TLC.

Open-access reader

About this research paper

What this paper is about

The standard plethysmographic method of measuring total lung capacity (TLC) has been reported to result in spuriously high estimates in patients with severe airway obstruction. The helium-dilution method is known to underestimate TLC in the same patients. To determine the magnitude of these possible errors we measured TLC by four methods in 20 patients with varying degrees of chronic obstructive lung disease and in 11 normal subjects. TLC was measured by (1) helium dilution (TLCHe); (2) a volume-displacement body plethysmograph, box volume being plotted against mouth pressure (TLCm); (3) the same body plethysmograph with volume plotted against pressure measured with an oesophageal balloon (TLCes); and (4) a radiological technique (TLCxr). In normal subjects there was no difference between TCLm (6.57 +/- 1.20) and TLCes (6.51 +/- 1.24). In the patients with chronic obstructive lung disease TLCm gave results significantly higher than those of any other method. If TLCes is taken as the closest estimate of true TLC, TLCm consistently overestimates and TLCHe underestimates TLC. There was no relationship between the degree of airway obstruction and (TLCm - TLCes) but there was between (TLCes - TLCHe) and severity of airway obstruction. We conclude that using mouth pressure in the plethysmographic measurement of TLC in patients with chronic obstructive lung disease results in consistent but slight overestimation of TLC.

Why it matters

OpenAlex reports 27 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The standard plethysmographic method of measuring total lung capacity (TLC) has been reported to result in spuriously high estimates in patients with severe airway obstruction. The helium-dilution method is known to underestimate TLC in the same patients. To determine the magnitude of these possible errors we measured TLC by four methods in 20 patients with varying degrees of chronic obstructive lung disease and in 11 normal subjects. TLC was measured by (1) helium dilution (TLCHe); (2) a volume-displacement body plethysmograph, box volume being plotted against mouth pressure (TLCm); (3) the same body plethysmograph with volume plotted against pressure measured with an oesophageal balloon (TLCes); and (4) a radiological technique (TLCxr). In normal subjects there was no difference between TCLm (6.57 +/- 1.20) and TLCes (6.51 +/- 1.24). In the patients with chronic obstructive lung disease TLCm gave results significantly higher than those of any other method. If TLCes is taken as the closest estimate of true TLC, TLCm consistently overestimates and TLCHe underestimates TLC. There was no relationship between the degree of airway obstruction and (TLCm - TLCes) but there was between (TLCes - TLCHe) and severity of airway obstruction. We conclude that using mouth pressure in the plethysmographic measurement of TLC in patients with chronic obstructive lung disease results in consistent but slight overestimation of TLC.

Key concepts: Plethysmograph, Medicine, Lung volumes, Obstructive lung disease, Airway obstruction, Lung, Restrictive lung disease, Airway

Related papers

Back to paper searchBrowse research topicsOriginal source
Errors in the measurement of total lung capacity in chronic obstructive lung disease. — Research Paper | ScholarLens