Evaluation of pulse oximetry in horses surgically treated for colic
Nora S. Matthews, Sandee M. Hartsfield, Elizabeth A. Sanders, Gwendolyn S. Light, Margaret Slater
Abstract
Nora S. Matthews, Sandee M. Hartsfield, Elizabeth A. Sanders, Gwendolyn S. Light, Margaret Slater
Abstract
All 43 horses anaesthetised for colic surgery were premedicated with xylazine or diazepam. Anaesthesia was induced with guaifenesin and ketamine, horses were placed in dorsal recumbency and anaesthesia was maintained with isoflurane in oxygen and mechanical ventilation. Haemoglobin saturation readings (SpO2) were taken with a pulse oximeter and compared with computed haemoglobin saturation (SaO2) from arterial blood samples. Readings were taken over a range of SaO2 of 78-100%, mean arterial blood pressure ranged from 24 to 108 mmHg and PaO2 ranged from 53 to 490 mmHg. Analysis of 107 readings showed that SpO2 values predicted SaO2 but time, blood pressure and individual horse did not. Correlation coefficients between SpO2 and SaO2 were 0.85 for all values and 0.88 for values at 30 min. Values for bias and precision were calculated for all SpO2 values and for readings separated into 3 saturation groups: normal, low normal, and abnormal. The pulse oximeter tended to underestimate SaO2 at all times, and was less precise as the saturation decreased.
OpenAlex reports 20 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
All 43 horses anaesthetised for colic surgery were premedicated with xylazine or diazepam. Anaesthesia was induced with guaifenesin and ketamine, horses were placed in dorsal recumbency and anaesthesia was maintained with isoflurane in oxygen and mechanical ventilation. Haemoglobin saturation readings (SpO2) were taken with a pulse oximeter and compared with computed haemoglobin saturation (SaO2) from arterial blood samples. Readings were taken over a range of SaO2 of 78-100%, mean arterial blood pressure ranged from 24 to 108 mmHg and PaO2 ranged from 53 to 490 mmHg. Analysis of 107 readings showed that SpO2 values predicted SaO2 but time, blood pressure and individual horse did not. Correlation coefficients between SpO2 and SaO2 were 0.85 for all values and 0.88 for values at 30 min. Values for bias and precision were calculated for all SpO2 values and for readings separated into 3 saturation groups: normal, low normal, and abnormal. The pulse oximeter tended to underestimate SaO2 at all times, and was less precise as the saturation decreased.
Key concepts: Xylazine, Pulse oximetry, Anesthesia, Isoflurane, Medicine, Ketamine, Blood pressure, Oxygen saturation