Review Article: Patient-Specific versus Standard Instrumentation for Total Knee Arthroplasty
Behnam Sharareh, Ran Schwarzkopf
Abstract
Behnam Sharareh, Ran Schwarzkopf
Abstract
This meta-analysis reviewed 12 studies comparing patient-specific instrumentation (PSI) with standard instrumentation (SI) for total knee arthroplasty in terms of postoperative coronal alignment and operation time. There is no significant difference between PSI and SI in terms of hip-knee-ankle angle (overall coronal alignment or mechanical axis), tibial coronal alignment, and operation time.
OpenAlex reports 26 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
This meta-analysis reviewed 12 studies comparing patient-specific instrumentation (PSI) with standard instrumentation (SI) for total knee arthroplasty in terms of postoperative coronal alignment and operation time. There is no significant difference between PSI and SI in terms of hip-knee-ankle angle (overall coronal alignment or mechanical axis), tibial coronal alignment, and operation time.
Key concepts: Coronal plane, Medicine, Total knee arthroplasty, Instrumentation (computer programming), Ankle, Arthroplasty, Surgery, Orthodontics