Breathing pattern and workload during automatic tube compensation, pressure support and T-piece trials in weaning patients

Ralf Kuhlen, M. Max, R. Dembinski, Sandra Terbeck, E. Jürgens, Rolf Rossaint

Open publisher page 34 citations

Abstract

BACKGROUND AND OBJECTIVE: Automatic tube compensation has been designed as a new ventilatory mode to compensate for the non-linear resistance of the endotracheal tube. The study investigated the effects of automatic tube compensation compared with breathing through a T-piece or pressure support during a trial of spontaneous breathing used for weaning patients from mechanical ventilation of the lungs. METHODS: Twelve patients were studied who were ready for weaning after prolonged mechanical ventilation (10.2 +/- 8.4 days) due to acute respiratory failure. Patients with chronic obstructive pulmonary disease were excluded. Thirty minutes of automatic tube compensation were compared with 30 min periods of 7 cmH2O pressure support and T-piece breathing. Breathing patterns and workload indices were measured at the end of each study period. RESULTS: During T-piece breathing, the peak inspiratory flow rate (0.65 +/- 0.20 L s(-1)) and minute ventilation (8.9 +/- 2.7L min(-1)) were lower than during either pressure support (peak inspiratory flow rate 0.81 +/- 0.25 L s(-1) minute ventilation 10.2 +/- 2.3 L min(-1), respectively) or automatic tube compensation (peak inspiratory flow rate 0.75 +/- 0.26L s(-1); minute ventilation 10.8 +/- 2.7 L min(-1)). The pressure-time product as well as patients' work of breathing were comparable during automatic tube compensation (pressure-time product 214.5 +/- 104.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)) and T-piece breathing (pressure-time product 208.3 +/- 121.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)), whereas pressure support resulted in a significant decrease in workload indices (pressure-time product 121.2 +/- 64.1 cmH2O s(-1) min(-1), patient work of breathing 0.7 +/- 0.4 J L(-1)). CONCLUSIONS: In weaning from mechanical lung ventilation, patients' work of breathing during spontaneous breathing trials is clearly reduced by the application of pressure support 7 cmH2O, whereas the workload during automatic tube compensation corresponded closely to the values during trials of breathing through a T-piece.

About this research paper

What this paper is about

BACKGROUND AND OBJECTIVE: Automatic tube compensation has been designed as a new ventilatory mode to compensate for the non-linear resistance of the endotracheal tube. The study investigated the effects of automatic tube compensation compared with breathing through a T-piece or pressure support during a trial of spontaneous breathing used for weaning patients from mechanical ventilation of the lungs. METHODS: Twelve patients were studied who were ready for weaning after prolonged mechanical ventilation (10.2 +/- 8.4 days) due to acute respiratory failure. Patients with chronic obstructive pulmonary disease were excluded. Thirty minutes of automatic tube compensation were compared with 30 min periods of 7 cmH2O pressure support and T-piece breathing. Breathing patterns and workload indices were measured at the end of each study period. RESULTS: During T-piece breathing, the peak inspiratory flow rate (0.65 +/- 0.20 L s(-1)) and minute ventilation (8.9 +/- 2.7L min(-1)) were lower than during either pressure support (peak inspiratory flow rate 0.81 +/- 0.25 L s(-1) minute ventilation 10.2 +/- 2.3 L min(-1), respectively) or automatic tube compensation (peak inspiratory flow rate 0.75 +/- 0.26L s(-1); minute ventilation 10.8 +/- 2.7 L min(-1)). The pressure-time product as well as patients' work of breathing were comparable during automatic tube compensation (pressure-time product 214.5 +/- 104.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)) and T-piece breathing (pressure-time product 208.3 +/- 121.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)), whereas pressure support resulted in a significant decrease in workload indices (pressure-time product 121.2 +/- 64.1 cmH2O s(-1) min(-1), patient work of breathing 0.7 +/- 0.4 J L(-1)). CONCLUSIONS: In weaning from mechanical lung ventilation, patients' work of breathing during spontaneous breathing trials is clearly reduced by the application of pressure support 7 cmH2O, whereas the workload during automatic tube compensation corresponded closely to the values during trials of breathing through a T-piece.

Why it matters

OpenAlex reports 34 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND AND OBJECTIVE: Automatic tube compensation has been designed as a new ventilatory mode to compensate for the non-linear resistance of the endotracheal tube. The study investigated the effects of automatic tube compensation compared with breathing through a T-piece or pressure support during a trial of spontaneous breathing used for weaning patients from mechanical ventilation of the lungs. METHODS: Twelve patients were studied who were ready for weaning after prolonged mechanical ventilation (10.2 +/- 8.4 days) due to acute respiratory failure. Patients with chronic obstructive pulmonary disease were excluded. Thirty minutes of automatic tube compensation were compared with 30 min periods of 7 cmH2O pressure support and T-piece breathing. Breathing patterns and workload indices were measured at the end of each study period. RESULTS: During T-piece breathing, the peak inspiratory flow rate (0.65 +/- 0.20 L s(-1)) and minute ventilation (8.9 +/- 2.7L min(-1)) were lower than during either pressure support (peak inspiratory flow rate 0.81 +/- 0.25 L s(-1) minute ventilation 10.2 +/- 2.3 L min(-1), respectively) or automatic tube compensation (peak inspiratory flow rate 0.75 +/- 0.26L s(-1); minute ventilation 10.8 +/- 2.7 L min(-1)). The pressure-time product as well as patients' work of breathing were comparable during automatic tube compensation (pressure-time product 214.5 +/- 104.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)) and T-piece breathing (pressure-time product 208.3 +/- 121.6 cmH2O s(-1) min(-1), patient work of breathing 1.1 +/- 0.4 J L(-1)), whereas pressure support resulted in a significant decrease in workload indices (pressure-time product 121.2 +/- 64.1 cmH2O s(-1) min(-1), patient work of breathing 0.7 +/- 0.4 J L(-1)). CONCLUSIONS: In weaning from mechanical lung ventilation, patients' work of breathing during spontaneous breathing trials is clearly reduced by the application of pressure support 7 cmH2O, whereas the workload during automatic tube compensation corresponded closely to the values during trials of breathing through a T-piece.

Key concepts: Medicine, Workload, Spontaneous breathing trial, Weaning, Breathing, Compensation (psychology), Anesthesia, Ventilator weaning

Related papers

Back to paper searchBrowse research topicsOriginal source
Breathing pattern and workload during automatic tube compensation, pressure support and T-piece trials in weaning patients — Research Paper | ScholarLens