2010Biochemia MedicaOpen access

Analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus

Jasna Juriček, Lovorka Đerek, Adriana Unić, Tihana Serdar, Domagoj Marijančević, Marcela Živković, Željko Romić

Open full text 5 citations

Abstract

Background: The objective of this study was to perform the analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus.The evaluation was performed according to the guidelines of the European Committee for Clinical Laboratory Standards (ECCLS). Materials and methods:The evaluation consisted of determination of within-run and between-run imprecision, inaccuracy and comparison with Olympus AU2700.The tested analytes were: glucose, creatinine, urate, total bilirubin, cholesterol, tryglicerides, calcium, phosphate, iron, unsaturated iron binding capacity, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, alpha-amylase, alkaline phosphatase, potassium, sodium, chloride, C-reactive protein, antistreptolysin O and rheumatoid factor.Results: The results showed low within-run and between-run imprecision and acceptable inaccuracy for all analytes.High correlation with AU2700 analyzer was found.Intercept and slope (with 95% confi dence interval) met the preferences of Passing-Bablok regression for all of the analytes except alanine aminotransferase, sodium, rheumatoid factor, creatine kinase, total bilirubin and tryglicerides.Conclusions: Olympus AU2700 plus analyzer shows acceptable precision and accuracy for majority of analytes, with the exception of sodium and chloride.Instrument is fully comparable with Olympus AU2700 analyzer for all but several analytes (ALT, RF, sodium, CK, total bilirubin and tryglicerides), where only some minor deviations (constant and proportional diff erence) were observed.

Open-access reader

About this research paper

What this paper is about

Background: The objective of this study was to perform the analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus.The evaluation was performed according to the guidelines of the European Committee for Clinical Laboratory Standards (ECCLS). Materials and methods:The evaluation consisted of determination of within-run and between-run imprecision, inaccuracy and comparison with Olympus AU2700.The tested analytes were: glucose, creatinine, urate, total bilirubin, cholesterol, tryglicerides, calcium, phosphate, iron, unsaturated iron binding capacity, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, alpha-amylase, alkaline phosphatase, potassium, sodium, chloride, C-reactive protein, antistreptolysin O and rheumatoid factor.Results: The results showed low within-run and between-run imprecision and acceptable inaccuracy for all analytes.High correlation with AU2700 analyzer was found.Intercept and slope (with 95% confi dence interval) met the preferences of Passing-Bablok regression for all of the analytes except alanine aminotransferase, sodium, rheumatoid factor, creatine kinase, total bilirubin and tryglicerides.Conclusions: Olympus AU2700 plus analyzer shows acceptable precision and accuracy for majority of analytes, with the exception of sodium and chloride.Instrument is fully comparable with Olympus AU2700 analyzer for all but several analytes (ALT, RF, sodium, CK, total bilirubin and tryglicerides), where only some minor deviations (constant and proportional diff erence) were observed.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Background: The objective of this study was to perform the analytical evaluation of the clinical chemistry analyzer Olympus AU2700 plus.The evaluation was performed according to the guidelines of the European Committee for Clinical Laboratory Standards (ECCLS). Materials and methods:The evaluation consisted of determination of within-run and between-run imprecision, inaccuracy and comparison with Olympus AU2700.The tested analytes were: glucose, creatinine, urate, total bilirubin, cholesterol, tryglicerides, calcium, phosphate, iron, unsaturated iron binding capacity, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, alpha-amylase, alkaline phosphatase, potassium, sodium, chloride, C-reactive protein, antistreptolysin O and rheumatoid factor.Results: The results showed low within-run and between-run imprecision and acceptable inaccuracy for all analytes.High correlation with AU2700 analyzer was found.Intercept and slope (with 95% confi dence interval) met the preferences of Passing-Bablok regression for all of the analytes except alanine aminotransferase, sodium, rheumatoid factor, creatine kinase, total bilirubin and tryglicerides.Conclusions: Olympus AU2700 plus analyzer shows acceptable precision and accuracy for majority of analytes, with the exception of sodium and chloride.Instrument is fully comparable with Olympus AU2700 analyzer for all but several analytes (ALT, RF, sodium, CK, total bilirubin and tryglicerides), where only some minor deviations (constant and proportional diff erence) were observed.

Key concepts: Spectrum analyzer, Medical physics, Analytical Chemistry (journal), Medicine, Chemistry, Engineering, Environmental chemistry, Telecommunications