2011American Journal of RoentgenologyOpen access

Comparison of Radiation Dose and Image Quality: 320-MDCT Versus 64-MDCT Coronary Angiography

Atif N. Khan, Faisal Khosa, Khurram Nasir, Aya Yassin, Melvin Eugene Clouse

Open full text 42 citations

Abstract

OBJECTIVE: The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating. MATERIALS AND METHODS: One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus. RESULTS: The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001). CONCLUSION: Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.

About this research paper

What this paper is about

OBJECTIVE: The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating. MATERIALS AND METHODS: One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus. RESULTS: The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001). CONCLUSION: Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.

Why it matters

OpenAlex reports 42 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: The purpose of this study was to compare radiation dose and image quality of 320- and 64-MDCT angiography using prospective gating. MATERIALS AND METHODS: One hundred seventy-four patients underwent 320-MDCT, and 95 patients underwent 64-MDCT. The scan parameters for 320-MDCT were 120 kVp, 400 mA, and gantry rotation of 350 milliseconds; the parameters for 64-MDCT were 120 kVp, 600 mA, and gantry rotation of 350 milliseconds. Effective dose (ED) was calculated from the dose-length product and a conversion factor (k = 0.014 mSv / mGy × cm). Two observers independently assessed image quality using a 3-point scale, where 3 denotes excellent quality and 1 denotes nondiagnostic quality, using a 16-segment model. Discrepancies were settled by consensus. RESULTS: The ED was significantly lower in patients undergoing 320-MDCT angiography, with a median ED of 4.4 mSv (interquartile range [IQR], 3.4-6.2 mSv), compared with 64-MDCT angiography, with a median ED of 6.2 mSv (IQR, 5.5-6.9 mSv) (p = 0.0001). In patients with a heart rate of 65 beats/min or less (92%), the median radiation dose using 320-MDCT was 4.1 mSv (IQR, 3.2-6.1 mSv), and that for 64-MDCT angiography was 6.2 mSv (IQR, 5.8-6.9 mSv) (p = 0.0001). In patients with heart rate greater than 65 beats/min (8%), the median dose was higher with 320-MDCT (8.7 mSv; IQR, 5.9-14.3 mSv) than with 64-MDCT (5.8 mSv; IQR, 5.3-6.7 mSv) (p = 0.02). Segmental image quality was significantly better for 320-MDCT (excellent or good quality, 96.66%; nondiagnostic quality, 0.1%) than for 64-MDCT angiography (excellent or good quality, 86%; nondiagnostic quality, 3.33%) (all p < 0.0001). CONCLUSION: Image quality was good for both 320- and 64-MDCT angiography. Overall radiation dose was significantly lower in 320-MDCT angiography when the heart rate was 65 beats/min or less. Every effort should be made to control heart rate to minimize radiation dose.

Key concepts: Medicine, Interquartile range, Nuclear medicine, Effective dose (radiation), Angiography, Radiation dose, Image quality, Radiology

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of Radiation Dose and Image Quality: 320-MDCT Versus 64-MDCT Coronary Angiography — Research Paper | ScholarLens