On the Adequacy of Half-value Layer as a Criterion of X-ray Quality in the Calibration of Dosemeters
N. M. Procter, J R Greening
Abstract
N. M. Procter, J R Greening
Abstract
Five thimble ionization chambers were compared with a free-air chamber using X rays produced at 20 to 55 kV with half-value layers between 0·05 and 0·20 mm Al. One chamber showed a difference of nearly 4 per cent in its correction factor for beams of the same H.V.L. produced at 20 and 55 kV. With this single exception H.V.L. appeared to be an adequate criterion of X-ray quality for the calibration of ionization chambers.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Five thimble ionization chambers were compared with a free-air chamber using X rays produced at 20 to 55 kV with half-value layers between 0·05 and 0·20 mm Al. One chamber showed a difference of nearly 4 per cent in its correction factor for beams of the same H.V.L. produced at 20 and 55 kV. With this single exception H.V.L. appeared to be an adequate criterion of X-ray quality for the calibration of ionization chambers.
Key concepts: Ionization chamber, Calibration, Ionization, X-ray, Nuclear medicine, Materials science, Half-value layer, Physics