Mammography radiation dose: initial results from Serbia based on mean glandular dose assessment for phantoms and patients
Olivera Ciraj‐Bjelac, S. Beciric, Danijela Arandjic, Duško Košutić, Maša Kovačević
Abstract
Open-access reader
Olivera Ciraj‐Bjelac, S. Beciric, Danijela Arandjic, Duško Košutić, Maša Kovačević
Abstract
Open-access reader
The primary objective of this study is to investigate the actual mammography dose levels in Serbia and establish a baseline dose database for the future screening programme. The mean glandular dose (MGD) was assessed for standard size breast substituted by 45 mm polymethyl methacrylate (PMMA) phantom and imaged under typical clinical conditions at 30 screen-film mammography facilities. Average MGD was (1.9 +/- 1.0) mGy, with a range of 0.12-5.2 mGy. These results were used for the calculation of the initial diagnostic reference levels in mammography in Serbia, which was set to 2.1 mGy for a standard breast. At some facilities, high doses were observed and the assessed MGD was supplemented by a patient dose survey. MGD was measured for 53 women having a diagnostic mammography in cranio-caudal (CC) and mediolateral oblique (MLO) projections at the largest mammography facilities in Serbia. The parameters recorded were: age, compressed breast thickness, tube potential (in kV), tube loading (in mAs) and the MLO projection angle. The average MGD per image was 2.8 mGy for the CC projection and 4.3 mGy for the MLO projection. Dose to breast was compared with dose for PMMA phantoms of various sizes, assessed at the same facilities. The results have indicated that phantom dose values can assist in setting optimisation activities in mammography. Both phantom and patient dose values indicated unnecessary high doses in facilities with an extremely high mammography workload. For these facilities, urgent dose-reduction measures and follow-up actions were recommended.
OpenAlex reports 17 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The primary objective of this study is to investigate the actual mammography dose levels in Serbia and establish a baseline dose database for the future screening programme. The mean glandular dose (MGD) was assessed for standard size breast substituted by 45 mm polymethyl methacrylate (PMMA) phantom and imaged under typical clinical conditions at 30 screen-film mammography facilities. Average MGD was (1.9 +/- 1.0) mGy, with a range of 0.12-5.2 mGy. These results were used for the calculation of the initial diagnostic reference levels in mammography in Serbia, which was set to 2.1 mGy for a standard breast. At some facilities, high doses were observed and the assessed MGD was supplemented by a patient dose survey. MGD was measured for 53 women having a diagnostic mammography in cranio-caudal (CC) and mediolateral oblique (MLO) projections at the largest mammography facilities in Serbia. The parameters recorded were: age, compressed breast thickness, tube potential (in kV), tube loading (in mAs) and the MLO projection angle. The average MGD per image was 2.8 mGy for the CC projection and 4.3 mGy for the MLO projection. Dose to breast was compared with dose for PMMA phantoms of various sizes, assessed at the same facilities. The results have indicated that phantom dose values can assist in setting optimisation activities in mammography. Both phantom and patient dose values indicated unnecessary high doses in facilities with an extremely high mammography workload. For these facilities, urgent dose-reduction measures and follow-up actions were recommended.
Key concepts: Mammography, Imaging phantom, Medicine, Polymethyl methacrylate, Nuclear medicine, Radiation dose, Dosimetry, Radiology