Evaluation of Interferences in Rate and Fixed-Time Nephelometric Assays of Specific Serum Proteins

Xavier Bossuyt, Norbert Blanckaert

Open full text 21 citations

Abstract

We performed interference studies for IgG, IgA, IgM, haptoglobin, and alpha1-antitrypsin assayed in serum, using either fixed-time nephelometry on the BN 100 from Behring or rate nephelometry on two analyzers from Beckman Instruments. For clear serum samples, results for IgG, IgA, IgM, and haptoglobin obtained with the three nephelometers showed good agreement. Values for alpha1-antitrypsin in clear sera were lower with the BN 100 than with the Array 360 or Immage. In lipemic samples, the BN 100 gave higher values than the Array 360 or Immage for all analytes except IgG. Addition of Intralipid to serum produced atypical reactions with the BN 100 (fixed-time nephelometry) but not with the Array 360 or Immage (rate nephelometry). The interference of lipemia on the BN 100 was also seen when the Beckman antibody was used, indicating that the effect was reagent-independent. For hemolyzed samples, the BN 100 gave higher values than the Array 360 or Immage for haptoglobin but not for the other analytes. Addition of increasing amounts of a hemolysate to serum revealed a negative interference in all assay systems. This effect was more pronounced with the Beckman reagent than with the Behring reagent in all three nephelometers and was independent of the type of instrument (fixed-time vs rate nephelometry).

Open-access reader

About this research paper

What this paper is about

We performed interference studies for IgG, IgA, IgM, haptoglobin, and alpha1-antitrypsin assayed in serum, using either fixed-time nephelometry on the BN 100 from Behring or rate nephelometry on two analyzers from Beckman Instruments. For clear serum samples, results for IgG, IgA, IgM, and haptoglobin obtained with the three nephelometers showed good agreement. Values for alpha1-antitrypsin in clear sera were lower with the BN 100 than with the Array 360 or Immage. In lipemic samples, the BN 100 gave higher values than the Array 360 or Immage for all analytes except IgG. Addition of Intralipid to serum produced atypical reactions with the BN 100 (fixed-time nephelometry) but not with the Array 360 or Immage (rate nephelometry). The interference of lipemia on the BN 100 was also seen when the Beckman antibody was used, indicating that the effect was reagent-independent. For hemolyzed samples, the BN 100 gave higher values than the Array 360 or Immage for haptoglobin but not for the other analytes. Addition of increasing amounts of a hemolysate to serum revealed a negative interference in all assay systems. This effect was more pronounced with the Beckman reagent than with the Behring reagent in all three nephelometers and was independent of the type of instrument (fixed-time vs rate nephelometry).

Why it matters

OpenAlex reports 21 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We performed interference studies for IgG, IgA, IgM, haptoglobin, and alpha1-antitrypsin assayed in serum, using either fixed-time nephelometry on the BN 100 from Behring or rate nephelometry on two analyzers from Beckman Instruments. For clear serum samples, results for IgG, IgA, IgM, and haptoglobin obtained with the three nephelometers showed good agreement. Values for alpha1-antitrypsin in clear sera were lower with the BN 100 than with the Array 360 or Immage. In lipemic samples, the BN 100 gave higher values than the Array 360 or Immage for all analytes except IgG. Addition of Intralipid to serum produced atypical reactions with the BN 100 (fixed-time nephelometry) but not with the Array 360 or Immage (rate nephelometry). The interference of lipemia on the BN 100 was also seen when the Beckman antibody was used, indicating that the effect was reagent-independent. For hemolyzed samples, the BN 100 gave higher values than the Array 360 or Immage for haptoglobin but not for the other analytes. Addition of increasing amounts of a hemolysate to serum revealed a negative interference in all assay systems. This effect was more pronounced with the Beckman reagent than with the Behring reagent in all three nephelometers and was independent of the type of instrument (fixed-time vs rate nephelometry).

Key concepts: Nephelometry, Haptoglobin, Chemistry, Analyte, Reagent, Chromatography, Antibody, Immunology

Related papers

Back to paper searchBrowse research topicsOriginal source
Evaluation of Interferences in Rate and Fixed-Time Nephelometric Assays of Specific Serum Proteins — Research Paper | ScholarLens