The Contribution of Vessel Volume Change and Blood Resistivity Change to the Electrical Impedance Pulse
T. M. Ravi Shankar, John G. Webster, Shu-Yong Shao
Abstract
T. M. Ravi Shankar, John G. Webster, Shu-Yong Shao
Abstract
An impedance pulse, recorded noninvasively, has contributions due to both the change in blood volume of the arteries and to the change in the blood resistivity. Other researchers have tried to quantify the relative contributions and have either underestimated or overestimated the contributions since they did not simulate the physiological conditions. We have used an in vitro flow circulation system to more closely simulate the physiological conditions and quantify the two contributions. We find that the blood resistivity change contribution is strong enough (21.5 percent of the arterial volume change contribution) to change the morphology of the impedance pulse. There is, however, a phase difference between the two contributions. As a result of this, the blood resistivity change contribution to the height of the impedance pulse will be less than 5.5 percent.
OpenAlex reports 44 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
An impedance pulse, recorded noninvasively, has contributions due to both the change in blood volume of the arteries and to the change in the blood resistivity. Other researchers have tried to quantify the relative contributions and have either underestimated or overestimated the contributions since they did not simulate the physiological conditions. We have used an in vitro flow circulation system to more closely simulate the physiological conditions and quantify the two contributions. We find that the blood resistivity change contribution is strong enough (21.5 percent of the arterial volume change contribution) to change the morphology of the impedance pulse. There is, however, a phase difference between the two contributions. As a result of this, the blood resistivity change contribution to the height of the impedance pulse will be less than 5.5 percent.
Key concepts: Electrical resistivity and conductivity, Electrical impedance, Biomedical engineering, Pulse (music), Blood flow, Materials science, Volume (thermodynamics), Blood volume