Unloading of the work of breathing by proportional assist ventilation in a lung model

Luca M. Bigatello, Masaji Nishimura, Hideaki Imanaka, Dean Hess, William R. Kimball, Robert M. Kacmarek

Open publisher page 31 citations

Abstract

OBJECTIVES: Proportional assist ventilation is devised to increase airway pressure in proportion to inspiratory effort. A systematic study of the performance of this new mode of ventilation has not been presented. We tested in the laboratory the capability of proportional assist ventilation to unload the work of breathing in proportion to ventilatory drive, under a variety of mechanical loads. DESIGN: During variations of "ventilatory drive" (i.e., tidal volume), unloading of the work of breathing by proportional assist ventilation was contrasted with unloading by pressure-support ventilation. SETTING: The respiratory laboratory of a university-affiliated teaching hospital. SUBJECT: A bellows-in-a-box lung model, powered by a sine wave air flow generator. INTERVENTIONS: Proportional assist and pressure-support ventilation were preset to provide comparable support at a baseline "ventilatory drive" of 0.7-L tidal volume. The set levels of proportional assist and pressure-support ventilation were subsequently applied to five tidal volumes, from 0.2 to 1.2 L. Three levels of inspiratory support and three settings of mechanical load were evaluated. MEASUREMENTS AND MAIN RESULTS: Proportional assist ventilation significantly (p < .05) reduced the work of breathing of the lung model at all but the lowest tidal volume (0.2 L). The preset proportion of ventilatory support (30%, 50%, and 70%) unloaded the work of breathing uniformly as ventilatory drive was varied at tidal volumes of > or = 0.5 L, but not always at tidal volumes of < or = 0.4 L. In contrast, pressure-support ventilation overassisted low tidal volumes and underassisted high tidal volumes (p < .05). CONCLUSIONS: In a lung model, a prototype system delivering proportional assist ventilation provided uniform unloading of the work of breathing as the ventilatory drive was varied within a tidal volume range of 0.5 to 1.2 L. These findings confirm the theoretical modeling of proportional assist ventilation. This system, however, failed to properly unload low tidal volumes of 0.2 to 0.4 L.

About this research paper

What this paper is about

OBJECTIVES: Proportional assist ventilation is devised to increase airway pressure in proportion to inspiratory effort. A systematic study of the performance of this new mode of ventilation has not been presented. We tested in the laboratory the capability of proportional assist ventilation to unload the work of breathing in proportion to ventilatory drive, under a variety of mechanical loads. DESIGN: During variations of "ventilatory drive" (i.e., tidal volume), unloading of the work of breathing by proportional assist ventilation was contrasted with unloading by pressure-support ventilation. SETTING: The respiratory laboratory of a university-affiliated teaching hospital. SUBJECT: A bellows-in-a-box lung model, powered by a sine wave air flow generator. INTERVENTIONS: Proportional assist and pressure-support ventilation were preset to provide comparable support at a baseline "ventilatory drive" of 0.7-L tidal volume. The set levels of proportional assist and pressure-support ventilation were subsequently applied to five tidal volumes, from 0.2 to 1.2 L. Three levels of inspiratory support and three settings of mechanical load were evaluated. MEASUREMENTS AND MAIN RESULTS: Proportional assist ventilation significantly (p < .05) reduced the work of breathing of the lung model at all but the lowest tidal volume (0.2 L). The preset proportion of ventilatory support (30%, 50%, and 70%) unloaded the work of breathing uniformly as ventilatory drive was varied at tidal volumes of > or = 0.5 L, but not always at tidal volumes of < or = 0.4 L. In contrast, pressure-support ventilation overassisted low tidal volumes and underassisted high tidal volumes (p < .05). CONCLUSIONS: In a lung model, a prototype system delivering proportional assist ventilation provided uniform unloading of the work of breathing as the ventilatory drive was varied within a tidal volume range of 0.5 to 1.2 L. These findings confirm the theoretical modeling of proportional assist ventilation. This system, however, failed to properly unload low tidal volumes of 0.2 to 0.4 L.

Why it matters

OpenAlex reports 31 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVES: Proportional assist ventilation is devised to increase airway pressure in proportion to inspiratory effort. A systematic study of the performance of this new mode of ventilation has not been presented. We tested in the laboratory the capability of proportional assist ventilation to unload the work of breathing in proportion to ventilatory drive, under a variety of mechanical loads. DESIGN: During variations of "ventilatory drive" (i.e., tidal volume), unloading of the work of breathing by proportional assist ventilation was contrasted with unloading by pressure-support ventilation. SETTING: The respiratory laboratory of a university-affiliated teaching hospital. SUBJECT: A bellows-in-a-box lung model, powered by a sine wave air flow generator. INTERVENTIONS: Proportional assist and pressure-support ventilation were preset to provide comparable support at a baseline "ventilatory drive" of 0.7-L tidal volume. The set levels of proportional assist and pressure-support ventilation were subsequently applied to five tidal volumes, from 0.2 to 1.2 L. Three levels of inspiratory support and three settings of mechanical load were evaluated. MEASUREMENTS AND MAIN RESULTS: Proportional assist ventilation significantly (p < .05) reduced the work of breathing of the lung model at all but the lowest tidal volume (0.2 L). The preset proportion of ventilatory support (30%, 50%, and 70%) unloaded the work of breathing uniformly as ventilatory drive was varied at tidal volumes of > or = 0.5 L, but not always at tidal volumes of < or = 0.4 L. In contrast, pressure-support ventilation overassisted low tidal volumes and underassisted high tidal volumes (p < .05). CONCLUSIONS: In a lung model, a prototype system delivering proportional assist ventilation provided uniform unloading of the work of breathing as the ventilatory drive was varied within a tidal volume range of 0.5 to 1.2 L. These findings confirm the theoretical modeling of proportional assist ventilation. This system, however, failed to properly unload low tidal volumes of 0.2 to 0.4 L.

Key concepts: Tidal volume, Ventilation (architecture), Work of breathing, Medicine, Pressure support ventilation, Respiratory minute volume, Artificial ventilation, Mechanical ventilation