Evaluation of image quality in computed radiography based mammography systems

Abhinav B. Singh, Vipin Bhwaria, Daniel J. Valentino

Open publisher page 0 citations

Abstract

Mammography is the most widely accepted procedure for the early detection of breast cancer and Computed Radiography (CR) is a cost-effective technology for digital mammography. We have demonstrated that CR mammography image quality is viable for Digital Mammography. The image quality of mammograms acquired using Computed Radiography technology was evaluated using the Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE). The measurements were made using a 28 kVp beam (RQA M-II) using 2 mm of Al as a filter and a target/filter combination of Mo/Mo. The acquired image bit depth was 16 bits and the pixel pitch for scanning was 50 microns. A Step-Wedge phantom (to measure the Contrast-to-noise ratio (CNR)) and the CDMAM 3.4 Contrast Detail phantom were also used to assess the image quality. The CNR values were observed at varying thickness of PMMA. The CDMAM 3.4 phantom results were plotted and compared to the EUREF acceptable and achievable values. The effect on image quality was measured using the physics metrics. A lower DQE was observed even with a higher MTF. This could be possibly due to a higher noise component present due to the way the scanner was configured. The CDMAM phantom scores demonstrated a contrast-detail comparable to the EUREF values. A cost-effective CR machine was optimized for high-resolution and high-contrast imaging.

About this research paper

What this paper is about

Mammography is the most widely accepted procedure for the early detection of breast cancer and Computed Radiography (CR) is a cost-effective technology for digital mammography. We have demonstrated that CR mammography image quality is viable for Digital Mammography. The image quality of mammograms acquired using Computed Radiography technology was evaluated using the Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE). The measurements were made using a 28 kVp beam (RQA M-II) using 2 mm of Al as a filter and a target/filter combination of Mo/Mo. The acquired image bit depth was 16 bits and the pixel pitch for scanning was 50 microns. A Step-Wedge phantom (to measure the Contrast-to-noise ratio (CNR)) and the CDMAM 3.4 Contrast Detail phantom were also used to assess the image quality. The CNR values were observed at varying thickness of PMMA. The CDMAM 3.4 phantom results were plotted and compared to the EUREF acceptable and achievable values. The effect on image quality was measured using the physics metrics. A lower DQE was observed even with a higher MTF. This could be possibly due to a higher noise component present due to the way the scanner was configured. The CDMAM phantom scores demonstrated a contrast-detail comparable to the EUREF values. A cost-effective CR machine was optimized for high-resolution and high-contrast imaging.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Mammography is the most widely accepted procedure for the early detection of breast cancer and Computed Radiography (CR) is a cost-effective technology for digital mammography. We have demonstrated that CR mammography image quality is viable for Digital Mammography. The image quality of mammograms acquired using Computed Radiography technology was evaluated using the Modulation Transfer Function (MTF), Noise Power Spectrum (NPS) and Detective Quantum Efficiency (DQE). The measurements were made using a 28 kVp beam (RQA M-II) using 2 mm of Al as a filter and a target/filter combination of Mo/Mo. The acquired image bit depth was 16 bits and the pixel pitch for scanning was 50 microns. A Step-Wedge phantom (to measure the Contrast-to-noise ratio (CNR)) and the CDMAM 3.4 Contrast Detail phantom were also used to assess the image quality. The CNR values were observed at varying thickness of PMMA. The CDMAM 3.4 phantom results were plotted and compared to the EUREF acceptable and achievable values. The effect on image quality was measured using the physics metrics. A lower DQE was observed even with a higher MTF. This could be possibly due to a higher noise component present due to the way the scanner was configured. The CDMAM phantom scores demonstrated a contrast-detail comparable to the EUREF values. A cost-effective CR machine was optimized for high-resolution and high-contrast imaging.

Key concepts: Detective quantum efficiency, Imaging phantom, Digital mammography, Mammography, Optical transfer function, Image quality, Computed radiography, Contrast-to-noise ratio

Related papers

Back to paper searchBrowse research topicsOriginal source
Evaluation of image quality in computed radiography based mammography systems — Research Paper | ScholarLens