Isotope Dilution Analysis of Bromate in Drinking Water Matrixes by Ion Chromatography with Inductively Coupled Plasma Mass Spectrometric Detection

John T. Creed, Carol A. Brockhoff

Open publisher page 38 citations

Abstract

Bromate is a disinfection byproduct in drinking water which is formed during the ozonation of source water containing bromide. This paper describes the analysis of bromate via ion chromatography-inductively coupled plasma mass spectrometry. The separation of bromate from interferences such as bromide and brominated haloacetic acids is achieved using a PA-100 column in combination with a 5 mM HNO3 + 25 mM NH4NO3 mobile phase. Polyatomic ions are observed on masses 79 and 81 in a synthetic phosphate matrix and in ozonated drinking waters. These polyatomic ions have been tentatively identified as PO3+ and H2PO3+. These polyatomic ions do not interfere with the detection of bromate because phosphate elutes prior to bromate. A polyatomic ion is observed on mass 81 in a synthetic sulfate matrix and in ozonated drinking waters. This polyatomic ion has been tentatively identified as HSO3+ and does not interfere with the detection of bromate because sulfate elutes after bromate. Isotope dilution analysis produces a relative standard deviation (RSD) of approximately 5% for both enriched isotopic additions at sample concentrations of 10 ng/g. The RSD associated with the direct analysis of bromate is 3.2% at sample concentrations of 10 ng/g. The bromate concentrations determined in ozonated drinking waters via isotope dilution analysis are within 10% of the concentrations determined via direct analysis for sample concentrations above 2 ng/g. The detection limit for the direct analysis of bromate via IC-ICPMS is 0.3 ng/g.

About this research paper

What this paper is about

Bromate is a disinfection byproduct in drinking water which is formed during the ozonation of source water containing bromide. This paper describes the analysis of bromate via ion chromatography-inductively coupled plasma mass spectrometry. The separation of bromate from interferences such as bromide and brominated haloacetic acids is achieved using a PA-100 column in combination with a 5 mM HNO3 + 25 mM NH4NO3 mobile phase. Polyatomic ions are observed on masses 79 and 81 in a synthetic phosphate matrix and in ozonated drinking waters. These polyatomic ions have been tentatively identified as PO3+ and H2PO3+. These polyatomic ions do not interfere with the detection of bromate because phosphate elutes prior to bromate. A polyatomic ion is observed on mass 81 in a synthetic sulfate matrix and in ozonated drinking waters. This polyatomic ion has been tentatively identified as HSO3+ and does not interfere with the detection of bromate because sulfate elutes after bromate. Isotope dilution analysis produces a relative standard deviation (RSD) of approximately 5% for both enriched isotopic additions at sample concentrations of 10 ng/g. The RSD associated with the direct analysis of bromate is 3.2% at sample concentrations of 10 ng/g. The bromate concentrations determined in ozonated drinking waters via isotope dilution analysis are within 10% of the concentrations determined via direct analysis for sample concentrations above 2 ng/g. The detection limit for the direct analysis of bromate via IC-ICPMS is 0.3 ng/g.

Why it matters

OpenAlex reports 38 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Bromate is a disinfection byproduct in drinking water which is formed during the ozonation of source water containing bromide. This paper describes the analysis of bromate via ion chromatography-inductively coupled plasma mass spectrometry. The separation of bromate from interferences such as bromide and brominated haloacetic acids is achieved using a PA-100 column in combination with a 5 mM HNO3 + 25 mM NH4NO3 mobile phase. Polyatomic ions are observed on masses 79 and 81 in a synthetic phosphate matrix and in ozonated drinking waters. These polyatomic ions have been tentatively identified as PO3+ and H2PO3+. These polyatomic ions do not interfere with the detection of bromate because phosphate elutes prior to bromate. A polyatomic ion is observed on mass 81 in a synthetic sulfate matrix and in ozonated drinking waters. This polyatomic ion has been tentatively identified as HSO3+ and does not interfere with the detection of bromate because sulfate elutes after bromate. Isotope dilution analysis produces a relative standard deviation (RSD) of approximately 5% for both enriched isotopic additions at sample concentrations of 10 ng/g. The RSD associated with the direct analysis of bromate is 3.2% at sample concentrations of 10 ng/g. The bromate concentrations determined in ozonated drinking waters via isotope dilution analysis are within 10% of the concentrations determined via direct analysis for sample concentrations above 2 ng/g. The detection limit for the direct analysis of bromate via IC-ICPMS is 0.3 ng/g.

Key concepts: Bromate, Chemistry, Polyatomic ion, Isotope dilution, Inductively coupled plasma mass spectrometry, Bromide, Mass spectrometry, Ion chromatography