2013Bangladesh Journal of Medical PhysicsOpen access

Recombination Parameters of some Fabricated Ionization Chambers

A. Nishara Begum, Nobuhisa Takata

Open full text 2 citations

Abstract

Some pancake and spherical type ionization chambers of various size have been designed and fabricated for absolute air kerma measurement in 60Co and 137Cs ?-ray fields at the Primary Standard Dosimetry Laboratory (PSDL), National Institute of Advanced Industrial Science & Technology (AIST), Tsukuba, Japan. Values of recombination parameters A and m2g of these ionization chambers are obtained using a method proposed by De Almeida and Niatel and adopted by Boutillon. For absolute air kerma measurement, it is important to obtain accurate signal currents by correcting it for ion losses using the values of A and m2g of each of the ionization chambers. It has been observed from the measurements that the recombination parameters for pancake ionization chamber are smaller than spherical ionization chambers and for spherical type ionization chamber recombination parameter values depend on the dimensions of the electrodes and also the size of the chambers. DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14694 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 101-106

Open-access reader

About this research paper

What this paper is about

Some pancake and spherical type ionization chambers of various size have been designed and fabricated for absolute air kerma measurement in 60Co and 137Cs ?-ray fields at the Primary Standard Dosimetry Laboratory (PSDL), National Institute of Advanced Industrial Science & Technology (AIST), Tsukuba, Japan. Values of recombination parameters A and m2g of these ionization chambers are obtained using a method proposed by De Almeida and Niatel and adopted by Boutillon. For absolute air kerma measurement, it is important to obtain accurate signal currents by correcting it for ion losses using the values of A and m2g of each of the ionization chambers. It has been observed from the measurements that the recombination parameters for pancake ionization chamber are smaller than spherical ionization chambers and for spherical type ionization chamber recombination parameter values depend on the dimensions of the electrodes and also the size of the chambers. DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14694 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 101-106

Why it matters

OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Some pancake and spherical type ionization chambers of various size have been designed and fabricated for absolute air kerma measurement in 60Co and 137Cs ?-ray fields at the Primary Standard Dosimetry Laboratory (PSDL), National Institute of Advanced Industrial Science & Technology (AIST), Tsukuba, Japan. Values of recombination parameters A and m2g of these ionization chambers are obtained using a method proposed by De Almeida and Niatel and adopted by Boutillon. For absolute air kerma measurement, it is important to obtain accurate signal currents by correcting it for ion losses using the values of A and m2g of each of the ionization chambers. It has been observed from the measurements that the recombination parameters for pancake ionization chamber are smaller than spherical ionization chambers and for spherical type ionization chamber recombination parameter values depend on the dimensions of the electrodes and also the size of the chambers. DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14694 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 101-106

Key concepts: Kerma, Ionization chamber, Ionization, Atomic physics, Ion, Dosimetry, Materials science, Recombination

Related papers

Back to paper searchBrowse research topicsOriginal source
Recombination Parameters of some Fabricated Ionization Chambers — Research Paper | ScholarLens