Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100

Róbert de Jonge, Rob Brouwer, Marcel van Rijn, B.A.C. van Acker, Henk J.A.M. Otten, Jan Lindemans

Open publisher page 37 citations

Abstract

BACKGROUND: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting. METHODS: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution. RESULTS: The automated WBC count (x10(6)/L) was highly correlated with that of the microscopic reference method (r(2)=0.95; WBC-analyzer=0.97 x WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r(2)=0.92; WBC-analyzer=0.99 x WBC-reference method+32; n=36), neutrophil count (r(2)=0.94; WBC-analyzer=0.91 x WBC-reference method+6; n=35), and monocyte count (r(2)=0.73; WBC-analyzer=0.83 x WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50 x 10(6)/L (coefficient of variation 20%). CONCLUSIONS: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.

About this research paper

What this paper is about

BACKGROUND: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting. METHODS: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution. RESULTS: The automated WBC count (x10(6)/L) was highly correlated with that of the microscopic reference method (r(2)=0.95; WBC-analyzer=0.97 x WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r(2)=0.92; WBC-analyzer=0.99 x WBC-reference method+32; n=36), neutrophil count (r(2)=0.94; WBC-analyzer=0.91 x WBC-reference method+6; n=35), and monocyte count (r(2)=0.73; WBC-analyzer=0.83 x WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50 x 10(6)/L (coefficient of variation 20%). CONCLUSIONS: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.

Why it matters

OpenAlex reports 37 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting. METHODS: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution. RESULTS: The automated WBC count (x10(6)/L) was highly correlated with that of the microscopic reference method (r(2)=0.95; WBC-analyzer=0.97 x WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r(2)=0.92; WBC-analyzer=0.99 x WBC-reference method+32; n=36), neutrophil count (r(2)=0.94; WBC-analyzer=0.91 x WBC-reference method+6; n=35), and monocyte count (r(2)=0.73; WBC-analyzer=0.83 x WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50 x 10(6)/L (coefficient of variation 20%). CONCLUSIONS: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.

Key concepts: Leukocyte Counts, Pleural fluid, Medicine, Pathology, Immunology, Pleural effusion, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100 — Research Paper | ScholarLens