Simultaneous determination of chromium( III ) and chromium( VI ) in aqueous solutions by ion chromatography and chemiluminescence detection
Bente Gammelgaard, Ole Jøns, Bent Eichstedt Nielsen
Abstract
Bente Gammelgaard, Ole Jøns, Bent Eichstedt Nielsen
Abstract
A method for the simultaneous determination of chromium(iii) and chromium(vi) in a flow system based on chemiluminescence was developed. A Dionex cation-exchange guard column was used to separate chromium(iii) from chromium(vi), and chromium(vi) was reduced by potassium sulfite, whereupon both species were detected by use of the luminol-hydrogen peroxide chemiluminescence system. Linear calibration for both species was established over the concentration range 1-1000 micrograms l-1. The precision at the 20 micrograms l-1 level was 3.5% for chromium(iii) and 3.3% for chromium(vi), respectively. The detection limit was 0.5 micrograms l-1 for both species. Data were in agreement with Zeeman-effect background corrected atomic absorption spectrometry measurements.
OpenAlex reports 62 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A method for the simultaneous determination of chromium(iii) and chromium(vi) in a flow system based on chemiluminescence was developed. A Dionex cation-exchange guard column was used to separate chromium(iii) from chromium(vi), and chromium(vi) was reduced by potassium sulfite, whereupon both species were detected by use of the luminol-hydrogen peroxide chemiluminescence system. Linear calibration for both species was established over the concentration range 1-1000 micrograms l-1. The precision at the 20 micrograms l-1 level was 3.5% for chromium(iii) and 3.3% for chromium(vi), respectively. The detection limit was 0.5 micrograms l-1 for both species. Data were in agreement with Zeeman-effect background corrected atomic absorption spectrometry measurements.
Key concepts: Chromium, Chemiluminescence, Chemistry, Detection limit, Chromatography, Hydrogen peroxide, Aqueous solution, Luminol