SU-D-BRE-04: Evaluating the Dose Accuracy of a 2D Ion Chamber Array in High Dose Rate Pencil Beam Scanning Proton Beam

Luis Augusto Perles, Anthony E. Mascia, F Piskulich, Richard Lepage, Yi Zhang, Annelise Giebeler, Lei Dong

Open publisher page 0 citations

Abstract

Purpose: To evaluate the absolute dose accuracy of the PTW Octavius 729 XDR 2D ion chamber array at a high dose rate pencil beam scanning proton therapy facility. Methods: A set of 18 plans were created in our treatment planning system, each of which comprising a unique combination of field sizes (FS), length of spread out of Bragg peaks (SOBP) and depths. The parameters used were: FS of 5×5cm2, 10×10cm2 and 15×15cm2; flat SOBP of 5cm and 10cm; and isocenter depths of 10cm, 15cm and 20cm, which coincides with the center of the SOBP. The 2D array detector was positioned at the machine isocenter and the appropriate amount of solid water was used to match the planned depths of 10, 15 and 20 cm water equivalent depth. Subsequently, we measured the absolute dose at isocenter using a CC04 ion chamber in a 1D water tank. Both 2D array and CC04 were previously cross calibrated. We also collected the MU rates used by our proton machine from the log files. Results: The relative differences between the CC04 and the 2D array can be summarized into two groups, one with 5 cm SOBP and another with 10 cm SOBP. Plotting these datasets against FS shows that the 2D array response for high dose rate fields (FS of 5×5cm2 and 5cm SOBP) can be up to 2% lower. Similarly, plotting them against isocenter depths reveals the detector's response can be up to 2% lower for higher energy beams (about 200MeV nominal). The MU rate found in the machine log files for 5cm SOBP's were as high as twice the MU rate for the 10cm SOBP. Conclusion: The 2D array dose response showed a dose rate effect in scanning pencil beam delivery, which needs to be corrected to achieve a better dose accuracy.

About this research paper

What this paper is about

Purpose: To evaluate the absolute dose accuracy of the PTW Octavius 729 XDR 2D ion chamber array at a high dose rate pencil beam scanning proton therapy facility. Methods: A set of 18 plans were created in our treatment planning system, each of which comprising a unique combination of field sizes (FS), length of spread out of Bragg peaks (SOBP) and depths. The parameters used were: FS of 5×5cm2, 10×10cm2 and 15×15cm2; flat SOBP of 5cm and 10cm; and isocenter depths of 10cm, 15cm and 20cm, which coincides with the center of the SOBP. The 2D array detector was positioned at the machine isocenter and the appropriate amount of solid water was used to match the planned depths of 10, 15 and 20 cm water equivalent depth. Subsequently, we measured the absolute dose at isocenter using a CC04 ion chamber in a 1D water tank. Both 2D array and CC04 were previously cross calibrated. We also collected the MU rates used by our proton machine from the log files. Results: The relative differences between the CC04 and the 2D array can be summarized into two groups, one with 5 cm SOBP and another with 10 cm SOBP. Plotting these datasets against FS shows that the 2D array response for high dose rate fields (FS of 5×5cm2 and 5cm SOBP) can be up to 2% lower. Similarly, plotting them against isocenter depths reveals the detector's response can be up to 2% lower for higher energy beams (about 200MeV nominal). The MU rate found in the machine log files for 5cm SOBP's were as high as twice the MU rate for the 10cm SOBP. Conclusion: The 2D array dose response showed a dose rate effect in scanning pencil beam delivery, which needs to be corrected to achieve a better dose accuracy.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: To evaluate the absolute dose accuracy of the PTW Octavius 729 XDR 2D ion chamber array at a high dose rate pencil beam scanning proton therapy facility. Methods: A set of 18 plans were created in our treatment planning system, each of which comprising a unique combination of field sizes (FS), length of spread out of Bragg peaks (SOBP) and depths. The parameters used were: FS of 5×5cm2, 10×10cm2 and 15×15cm2; flat SOBP of 5cm and 10cm; and isocenter depths of 10cm, 15cm and 20cm, which coincides with the center of the SOBP. The 2D array detector was positioned at the machine isocenter and the appropriate amount of solid water was used to match the planned depths of 10, 15 and 20 cm water equivalent depth. Subsequently, we measured the absolute dose at isocenter using a CC04 ion chamber in a 1D water tank. Both 2D array and CC04 were previously cross calibrated. We also collected the MU rates used by our proton machine from the log files. Results: The relative differences between the CC04 and the 2D array can be summarized into two groups, one with 5 cm SOBP and another with 10 cm SOBP. Plotting these datasets against FS shows that the 2D array response for high dose rate fields (FS of 5×5cm2 and 5cm SOBP) can be up to 2% lower. Similarly, plotting them against isocenter depths reveals the detector's response can be up to 2% lower for higher energy beams (about 200MeV nominal). The MU rate found in the machine log files for 5cm SOBP's were as high as twice the MU rate for the 10cm SOBP. Conclusion: The 2D array dose response showed a dose rate effect in scanning pencil beam delivery, which needs to be corrected to achieve a better dose accuracy.

Key concepts: Sobp, Isocenter, Pencil-beam scanning, Ionization chamber, Bragg peak, Pencil (optics), Proton therapy, Proton

Related papers

Back to paper searchBrowse research topicsOriginal source
SU-D-BRE-04: Evaluating the Dose Accuracy of a 2D Ion Chamber Array in High Dose Rate Pencil Beam Scanning Proton Beam — Research Paper | ScholarLens