Field-testing and validation in a primary care setting of a point-of-care test for C-reactive protein
David Seamark, Susan Backhouse, Roy Powell
Abstract
David Seamark, Susan Backhouse, Roy Powell
Abstract
BACKGROUND: Use of point-of-care testing (POCT) in primary care has increased. There is a need for high-quality field evaluation of POCT before deployment can be considered. METHOD: A POCT system for C-reactive protein was evaluated in a routine general practice setting. The standard laboratory method was a dry slide method based in a routine hospital laboratory. RESULTS: Scatterplots for both venous and capillary blood POCT system results versus the standard laboratory assay produced correlation coefficients of greater than 0.96. Bland-Altman plots indicated that 95% of venous and capillary POCT results fell within +/-10 mg/L of the mean value with no clinically significant difference from laboratory results. CONCLUSIONS: The POCT system performed reliably in a routine general practice setting with satisfactory performance against an accepted laboratory method.
OpenAlex reports 42 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
BACKGROUND: Use of point-of-care testing (POCT) in primary care has increased. There is a need for high-quality field evaluation of POCT before deployment can be considered. METHOD: A POCT system for C-reactive protein was evaluated in a routine general practice setting. The standard laboratory method was a dry slide method based in a routine hospital laboratory. RESULTS: Scatterplots for both venous and capillary blood POCT system results versus the standard laboratory assay produced correlation coefficients of greater than 0.96. Bland-Altman plots indicated that 95% of venous and capillary POCT results fell within +/-10 mg/L of the mean value with no clinically significant difference from laboratory results. CONCLUSIONS: The POCT system performed reliably in a routine general practice setting with satisfactory performance against an accepted laboratory method.
Key concepts: Point-of-care testing, Point of care, Medicine, Medical laboratory, Laboratory test, External quality assessment, Pathology, Biochemical engineering