Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry
Ömer Faruk Recep, Hikmet Hasιripi, Nurullah Çağιl, Hikmet Sarιkatipoğlu
Abstract
Ömer Faruk Recep, Hikmet Hasιripi, Nurullah Çağιl, Hikmet Sarιkatipoğlu
Abstract
PURPOSE: To assess whether there is a significant difference in intraocular pressure (IOP) measurements between noncontact and applanation methods and if so, whether the difference is correlated with corneal thickness. SETTING: Işik Eye Clinic, Ankara, Turkey. METHODS: This prospective study comprised 120 eyes of 60 randomly selected patients. Noncontact tonometry, applanation tonometry, and pachymetry were performed in each eye. The results were compared and an analysis was performed to determine whether there was a correlation between the differences in IOP measurements and corneal thickness in the entire group and in subgroups composed of the halves and quarters of the group. RESULTS: The mean IOP in the entire group was 20.38 mm Hg +/- 4.97 (SD) with the noncontact tonometer and 18.84 +/- 4.47 mm Hg with the Goldmann applanation tonometer. The difference between the measurements with the noncontact method and those with the applanation method was statistically significant (P <.01) except in cases having a corneal thickness between 513 microm and 539 microm (P >.01). There was a positive correlation between corneal thickness and the difference in measurements by noncontact and applanation tonometry, and the magnitude of correlation was greater in cases with thicker corneas. CONCLUSION: The reliability of tonometers decreased with increasing corneal thickness, in which case higher readings were found, especially with the noncontact tonometer.
OpenAlex reports 63 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
PURPOSE: To assess whether there is a significant difference in intraocular pressure (IOP) measurements between noncontact and applanation methods and if so, whether the difference is correlated with corneal thickness. SETTING: Işik Eye Clinic, Ankara, Turkey. METHODS: This prospective study comprised 120 eyes of 60 randomly selected patients. Noncontact tonometry, applanation tonometry, and pachymetry were performed in each eye. The results were compared and an analysis was performed to determine whether there was a correlation between the differences in IOP measurements and corneal thickness in the entire group and in subgroups composed of the halves and quarters of the group. RESULTS: The mean IOP in the entire group was 20.38 mm Hg +/- 4.97 (SD) with the noncontact tonometer and 18.84 +/- 4.47 mm Hg with the Goldmann applanation tonometer. The difference between the measurements with the noncontact method and those with the applanation method was statistically significant (P <.01) except in cases having a corneal thickness between 513 microm and 539 microm (P >.01). There was a positive correlation between corneal thickness and the difference in measurements by noncontact and applanation tonometry, and the magnitude of correlation was greater in cases with thicker corneas. CONCLUSION: The reliability of tonometers decreased with increasing corneal thickness, in which case higher readings were found, especially with the noncontact tonometer.
Key concepts: Applanation tonometry, Intraocular pressure, Medicine, Ophthalmology, Goldmann Applanation Tonometer, Significant difference, Cornea, Mean difference