Performance characterization of computed radiography based mammography systems

Abhinav Singh, Nikunj Desai, Daniel J. Valentino

Open publisher page 3 citations

Abstract

Computed Radiography (CR) is a cost-effective technology for digital mammography. In order to optimize the quality of images obtained using CR Mammography, we characterized the effect on image quality of the electrooptical components of the CR imaging chain. The metrics used to assess the image quality included the Contrast to Noise Ratio (CNR), Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), Detective Quantum Efficiency (DQE) and Contrast Detail Response Phantom (CDMAM 3.4 Artinis Medical Systems). An 18×24 cm high-resolution granular phosphor imaging plate (AGFA MM3.0) was used to acquire the images. Contrast detail was measured using a GUI developed for the CDMAM phantom that was scored by independent observers. The range of theoretically acceptable values measured for the CR laser was (5-36) mW and voltage range for PMT's was (4-8) V. The light detection amplifier was investigated, and the optimal Laser Power and PMT gain used for scanning was measured. The tools that we used (CNR, MTF, NPS, DQE and Contrast-detail phantom) provided an effective means of selecting optimal values for the electro-optical components of the system. The procedure enabled us to obtain good quality CR mammograms that have less noise and improved contrast.

About this research paper

What this paper is about

Computed Radiography (CR) is a cost-effective technology for digital mammography. In order to optimize the quality of images obtained using CR Mammography, we characterized the effect on image quality of the electrooptical components of the CR imaging chain. The metrics used to assess the image quality included the Contrast to Noise Ratio (CNR), Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), Detective Quantum Efficiency (DQE) and Contrast Detail Response Phantom (CDMAM 3.4 Artinis Medical Systems). An 18×24 cm high-resolution granular phosphor imaging plate (AGFA MM3.0) was used to acquire the images. Contrast detail was measured using a GUI developed for the CDMAM phantom that was scored by independent observers. The range of theoretically acceptable values measured for the CR laser was (5-36) mW and voltage range for PMT's was (4-8) V. The light detection amplifier was investigated, and the optimal Laser Power and PMT gain used for scanning was measured. The tools that we used (CNR, MTF, NPS, DQE and Contrast-detail phantom) provided an effective means of selecting optimal values for the electro-optical components of the system. The procedure enabled us to obtain good quality CR mammograms that have less noise and improved contrast.

Why it matters

OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Computed Radiography (CR) is a cost-effective technology for digital mammography. In order to optimize the quality of images obtained using CR Mammography, we characterized the effect on image quality of the electrooptical components of the CR imaging chain. The metrics used to assess the image quality included the Contrast to Noise Ratio (CNR), Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), Detective Quantum Efficiency (DQE) and Contrast Detail Response Phantom (CDMAM 3.4 Artinis Medical Systems). An 18×24 cm high-resolution granular phosphor imaging plate (AGFA MM3.0) was used to acquire the images. Contrast detail was measured using a GUI developed for the CDMAM phantom that was scored by independent observers. The range of theoretically acceptable values measured for the CR laser was (5-36) mW and voltage range for PMT's was (4-8) V. The light detection amplifier was investigated, and the optimal Laser Power and PMT gain used for scanning was measured. The tools that we used (CNR, MTF, NPS, DQE and Contrast-detail phantom) provided an effective means of selecting optimal values for the electro-optical components of the system. The procedure enabled us to obtain good quality CR mammograms that have less noise and improved contrast.

Key concepts: Detective quantum efficiency, Imaging phantom, Computed radiography, Mammography, Optical transfer function, Digital mammography, Image quality, Optics

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance characterization of computed radiography based mammography systems — Research Paper | ScholarLens