Performance characterization of computed radiography based mammography systems
Abhinav Singh, Nikunj Desai, Daniel J. Valentino
Abstract
Abhinav Singh, Nikunj Desai, Daniel J. Valentino
Abstract
Computed Radiography (CR) is a cost-effective technology for digital mammography. In order to optimize the quality of images obtained using CR Mammography, we characterized the effect on image quality of the electrooptical components of the CR imaging chain. The metrics used to assess the image quality included the Contrast to Noise Ratio (CNR), Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), Detective Quantum Efficiency (DQE) and Contrast Detail Response Phantom (CDMAM 3.4 Artinis Medical Systems). An 18×24 cm high-resolution granular phosphor imaging plate (AGFA MM3.0) was used to acquire the images. Contrast detail was measured using a GUI developed for the CDMAM phantom that was scored by independent observers. The range of theoretically acceptable values measured for the CR laser was (5-36) mW and voltage range for PMT's was (4-8) V. The light detection amplifier was investigated, and the optimal Laser Power and PMT gain used for scanning was measured. The tools that we used (CNR, MTF, NPS, DQE and Contrast-detail phantom) provided an effective means of selecting optimal values for the electro-optical components of the system. The procedure enabled us to obtain good quality CR mammograms that have less noise and improved contrast.
OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Computed Radiography (CR) is a cost-effective technology for digital mammography. In order to optimize the quality of images obtained using CR Mammography, we characterized the effect on image quality of the electrooptical components of the CR imaging chain. The metrics used to assess the image quality included the Contrast to Noise Ratio (CNR), Modulation Transfer Function (MTF), Noise Power Spectrum (NPS), Detective Quantum Efficiency (DQE) and Contrast Detail Response Phantom (CDMAM 3.4 Artinis Medical Systems). An 18×24 cm high-resolution granular phosphor imaging plate (AGFA MM3.0) was used to acquire the images. Contrast detail was measured using a GUI developed for the CDMAM phantom that was scored by independent observers. The range of theoretically acceptable values measured for the CR laser was (5-36) mW and voltage range for PMT's was (4-8) V. The light detection amplifier was investigated, and the optimal Laser Power and PMT gain used for scanning was measured. The tools that we used (CNR, MTF, NPS, DQE and Contrast-detail phantom) provided an effective means of selecting optimal values for the electro-optical components of the system. The procedure enabled us to obtain good quality CR mammograms that have less noise and improved contrast.
Key concepts: Detective quantum efficiency, Imaging phantom, Computed radiography, Mammography, Optical transfer function, Digital mammography, Image quality, Optics