Evaluation of High-Field Orbitrap Fourier Transform Mass Spectrometer for Petroleomics
Konstantin O. Zhurov, Anton N. Kozhinov, Yury O. Tsybin
Abstract
Konstantin O. Zhurov, Anton N. Kozhinov, Yury O. Tsybin
Abstract
The performance of the high-field Orbitrap Fourier transform mass spectrometer (FTMS) is evaluated in the context of petroleum sample analysis. Pertinent characteristics including resolving power, mass accuracy, and spectral dynamic range are tested with resin and maltene samples. Comparisons are made to the latest requirements for petroleum analysis by Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS). The resolving power of the high-field compact Orbitrap mass analyzer (with a commercially provided resolution setting of 480,000 at 400 m / z ) is found to be sufficient for adequate characterization of the oil fractions under consideration. We demonstrate that the high-field Orbitrap FTMS may provide acceptable mass accuracy, comparable to FT-ICR MS, when an in-house developed recalibration procedure is employed. Presented data show that the high-field Orbitrap FTMS is particularly suitable for the study of lighter oil fractions, preferably with low sulfur content. Comprehensive analysis of more complex crude oil samples requires a further, at least 2-fold, increase in resolving power.
OpenAlex reports 81 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The performance of the high-field Orbitrap Fourier transform mass spectrometer (FTMS) is evaluated in the context of petroleum sample analysis. Pertinent characteristics including resolving power, mass accuracy, and spectral dynamic range are tested with resin and maltene samples. Comparisons are made to the latest requirements for petroleum analysis by Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS). The resolving power of the high-field compact Orbitrap mass analyzer (with a commercially provided resolution setting of 480,000 at 400 m / z ) is found to be sufficient for adequate characterization of the oil fractions under consideration. We demonstrate that the high-field Orbitrap FTMS may provide acceptable mass accuracy, comparable to FT-ICR MS, when an in-house developed recalibration procedure is employed. Presented data show that the high-field Orbitrap FTMS is particularly suitable for the study of lighter oil fractions, preferably with low sulfur content. Comprehensive analysis of more complex crude oil samples requires a further, at least 2-fold, increase in resolving power.
Key concepts: Orbitrap, Fourier transform ion cyclotron resonance, Mass spectrometry, Chemistry, Analytical Chemistry (journal), Fourier transform, Context (archaeology), Hybrid mass spectrometer