Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass Spectrometer

Alexander Makarov, Eduard Denisov, Alexander Kholomeev, Wilko Balschun, Oliver Lange, Kerstin Strupat, Stevan Horning

Open publisher page 765 citations

Abstract

Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60,000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100,000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.

About this research paper

What this paper is about

Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60,000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100,000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.

Why it matters

OpenAlex reports 765 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Design and performance of a novel hybrid mass spectrometer is described. It couples a linear ion trap mass spectrometer to an orbitrap mass analyzer via an rf-only trapping quadrupole with a curved axis. The latter injects pulsed ion beams into a rapidly changing electric field in the orbitrap wherein they are trapped at high kinetic energies around an inner electrode. Image current detection is subsequently performed after a stable electrostatic field is achieved. Fourier transformation of the acquired transient allows wide mass range detection with high resolving power, mass accuracy, and dynamic range. The entire instrument operates in LC/MS mode (1 spectrum/s) with nominal mass resolving power of 60,000 and uses automatic gain control to provide high-accuracy mass measurements, within 2 ppm using internal standards and within 5 ppm with external calibration. The maximum resolving power exceeds 100,000 (fwhm). Rapid, automated data-dependent capabilities enable real-time acquisition of up to three high-mass accuracy MS/MS spectra per second.

Key concepts: Chemistry, Orbitrap, Quadrupole ion trap, Ion trap, Mass spectrometry, Hybrid mass spectrometer, Ion, Trap (plumbing)