Spiral exchange cannula for the occluded drainage catheter.
A. H. McCain, Janice L. Vucinich, Jan F. Hawkins, Irvin F. Hawkins
Abstract
A. H. McCain, Janice L. Vucinich, Jan F. Hawkins, Irvin F. Hawkins
Abstract
A technique for exchange of occluded drainage catheters or placement of two guide wires into a target using a spiral exchange cannula is described. A spiral exchange cannula with a preloaded sheath and threads at the distal end is "screwed" into the drainage catheter, and the sheath is advanced over the catheter. This method prevents dislodgment of the drainage catheter and also permits easy catheter exchange when the lumen of the drainage catheter is occluded.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A technique for exchange of occluded drainage catheters or placement of two guide wires into a target using a spiral exchange cannula is described. A spiral exchange cannula with a preloaded sheath and threads at the distal end is "screwed" into the drainage catheter, and the sheath is advanced over the catheter. This method prevents dislodgment of the drainage catheter and also permits easy catheter exchange when the lumen of the drainage catheter is occluded.
Key concepts: Medicine, Cannula, Catheter, Drainage, Spiral (railway), Lumen (anatomy), Surgery, Mechanical engineering