2013International Journal of Medical Physics Clinical Engineering and Radiation OncologyOpen access

Dosimetric Comparison of Craniospinal Axis Irradiation (CSI) Treatments Using Helical Tomotherapy, SmartarcTM, and 3D Conventional Radiation Therapy

Pamela Myers, S Stathakis, Alonso N. Gutiérrez, C Esquivel, Panayiotis Mavroidis, Niko Papanikolaou

Open full text 18 citations

Abstract

Purpose: Craniospinal axis irradiation (CSI) is a method of treating various central nervous system malignancies. The large target volume typically includes entire spinal cord and whole brain. Dosimetric comparison was performed between tomotherapy, volumetric modulated arc therapy (VMAT), and 3D conformal radiation therapy (3D-CRT) for CSI. Methods and Materials: Five (n = 5) CSI patients were planned using 3D-CRT, VMAT, and tomotherapy (normalized such that 95% of PTV received at least 23.4 Gy in 13 fractions). Plans were compared using PTV conformity number (CN) and homogeneity index (HI), normal tissue (NT) dose statistics, integral dose, and treatment time. Results: On average, tomotherapy plans showed higher CN (0.932 vs. 0.860 and 0.672 for SmartArc and 3D-CRT). In terms of HI, VMAT plans consistently showed better dose homogeneity (1.07 vs. 1.15 and 1.13 for tomotherapy and 3D-CRT). SmartArc delivered lower maximum dose for majority of NT, but higher mean dose. 3D-CRT plans delivered higher maximum dose but lower mean dose to NT. Conclusions: SmartArc treatments achieved better PTV homogeneity and reduced maximum dose to NT. Tomotherapy showed better target conformity, but 3D-CRT was shown to reduce mean dose to NT. Integral doses were similar between treatment modalities, but tomotherapy treatment times were much longer.

Open-access reader

About this research paper

What this paper is about

Purpose: Craniospinal axis irradiation (CSI) is a method of treating various central nervous system malignancies. The large target volume typically includes entire spinal cord and whole brain. Dosimetric comparison was performed between tomotherapy, volumetric modulated arc therapy (VMAT), and 3D conformal radiation therapy (3D-CRT) for CSI. Methods and Materials: Five (n = 5) CSI patients were planned using 3D-CRT, VMAT, and tomotherapy (normalized such that 95% of PTV received at least 23.4 Gy in 13 fractions). Plans were compared using PTV conformity number (CN) and homogeneity index (HI), normal tissue (NT) dose statistics, integral dose, and treatment time. Results: On average, tomotherapy plans showed higher CN (0.932 vs. 0.860 and 0.672 for SmartArc and 3D-CRT). In terms of HI, VMAT plans consistently showed better dose homogeneity (1.07 vs. 1.15 and 1.13 for tomotherapy and 3D-CRT). SmartArc delivered lower maximum dose for majority of NT, but higher mean dose. 3D-CRT plans delivered higher maximum dose but lower mean dose to NT. Conclusions: SmartArc treatments achieved better PTV homogeneity and reduced maximum dose to NT. Tomotherapy showed better target conformity, but 3D-CRT was shown to reduce mean dose to NT. Integral doses were similar between treatment modalities, but tomotherapy treatment times were much longer.

Why it matters

OpenAlex reports 18 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: Craniospinal axis irradiation (CSI) is a method of treating various central nervous system malignancies. The large target volume typically includes entire spinal cord and whole brain. Dosimetric comparison was performed between tomotherapy, volumetric modulated arc therapy (VMAT), and 3D conformal radiation therapy (3D-CRT) for CSI. Methods and Materials: Five (n = 5) CSI patients were planned using 3D-CRT, VMAT, and tomotherapy (normalized such that 95% of PTV received at least 23.4 Gy in 13 fractions). Plans were compared using PTV conformity number (CN) and homogeneity index (HI), normal tissue (NT) dose statistics, integral dose, and treatment time. Results: On average, tomotherapy plans showed higher CN (0.932 vs. 0.860 and 0.672 for SmartArc and 3D-CRT). In terms of HI, VMAT plans consistently showed better dose homogeneity (1.07 vs. 1.15 and 1.13 for tomotherapy and 3D-CRT). SmartArc delivered lower maximum dose for majority of NT, but higher mean dose. 3D-CRT plans delivered higher maximum dose but lower mean dose to NT. Conclusions: SmartArc treatments achieved better PTV homogeneity and reduced maximum dose to NT. Tomotherapy showed better target conformity, but 3D-CRT was shown to reduce mean dose to NT. Integral doses were similar between treatment modalities, but tomotherapy treatment times were much longer.

Key concepts: Tomotherapy, Nuclear medicine, Medicine, Radiation therapy, Radiology

Related papers

Back to paper searchBrowse research topicsOriginal source
Dosimetric Comparison of Craniospinal Axis Irradiation (CSI) Treatments Using Helical Tomotherapy, SmartarcTM, and 3D Conventional Radiation Therapy — Research Paper | ScholarLens