SU-E-T-535: Analysis of Results from an Anthropomorphic Stereotactic Radiosurgery Phantom

C Amador, A Molineu, Susan A. Smith, Nadia Hernandez, D Followill, Marilyn Stovall

Open publisher page 0 citations

Abstract

Purpose: To analyze 634 irradiations of the stereotactic radiosurgery (SRS) anthropomorphic phantom. Method and Materials:The anthropomorphic SRS head phantom consists of a water-fillable head shaped plastic shell that has been altered to accept imaging and dosimetry inserts. The water-fillable imaging insert houses a 1.9 cm diameter nylon sphere that allows localization with either CT or MRI. The dosimetry insert houses TLD and radiochromic film. Institutions imaged, planned and treated the phantom. Plans were designed to deliver 30 Gy to the center of the target that was covered by an isodose line typically between 50% and 85%. The treated volume, assumed to be an ellipsoid defined by the prescription dose, was calculated from film profiles and compared to distances reported by the institution. Results: Between 2000 and 2010, 125 of the irradiations were performed with Gamma Knife and 509 were performed with linear accelerators. Most of these irradiations were performed as a method of independent verification of treatment delivery. Gamma Knife TLD and treatment volume results were on average 0.98±0.03 and 1.09±0.11, respectively. Linear accelerator TLD and treatment volume results were on average 1.00±0.03 and 1.04±0.15, respectively. The percentage of institutions, using accelerators, meeting the QA guidelines for dose to the target, treated volume, ratio of measured treated volume to target volume and minimum dose to target were 93%, 90%, 80%, and 80%, respectively. For Gamma Knifes the corresponding percentages were 91%, 98%, 88%, and 49%. The percent of institutions meeting all four criteria were 54% and 39% for accelerators and Gamma Knife, respectively. Conclusion: Though most SRS phantom irradiations met several of the individual criteria, only 50% met all four. Failure to plan adequate target coverage and/or deliver adequate target coverage contributed to the low percentages for minimum dose to target compliance. Work was supported by PHS grants CA10953 and CA81647 awarded by the NCI, DHHS.

About this research paper

What this paper is about

Purpose: To analyze 634 irradiations of the stereotactic radiosurgery (SRS) anthropomorphic phantom. Method and Materials:The anthropomorphic SRS head phantom consists of a water-fillable head shaped plastic shell that has been altered to accept imaging and dosimetry inserts. The water-fillable imaging insert houses a 1.9 cm diameter nylon sphere that allows localization with either CT or MRI. The dosimetry insert houses TLD and radiochromic film. Institutions imaged, planned and treated the phantom. Plans were designed to deliver 30 Gy to the center of the target that was covered by an isodose line typically between 50% and 85%. The treated volume, assumed to be an ellipsoid defined by the prescription dose, was calculated from film profiles and compared to distances reported by the institution. Results: Between 2000 and 2010, 125 of the irradiations were performed with Gamma Knife and 509 were performed with linear accelerators. Most of these irradiations were performed as a method of independent verification of treatment delivery. Gamma Knife TLD and treatment volume results were on average 0.98±0.03 and 1.09±0.11, respectively. Linear accelerator TLD and treatment volume results were on average 1.00±0.03 and 1.04±0.15, respectively. The percentage of institutions, using accelerators, meeting the QA guidelines for dose to the target, treated volume, ratio of measured treated volume to target volume and minimum dose to target were 93%, 90%, 80%, and 80%, respectively. For Gamma Knifes the corresponding percentages were 91%, 98%, 88%, and 49%. The percent of institutions meeting all four criteria were 54% and 39% for accelerators and Gamma Knife, respectively. Conclusion: Though most SRS phantom irradiations met several of the individual criteria, only 50% met all four. Failure to plan adequate target coverage and/or deliver adequate target coverage contributed to the low percentages for minimum dose to target compliance. Work was supported by PHS grants CA10953 and CA81647 awarded by the NCI, DHHS.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: To analyze 634 irradiations of the stereotactic radiosurgery (SRS) anthropomorphic phantom. Method and Materials:The anthropomorphic SRS head phantom consists of a water-fillable head shaped plastic shell that has been altered to accept imaging and dosimetry inserts. The water-fillable imaging insert houses a 1.9 cm diameter nylon sphere that allows localization with either CT or MRI. The dosimetry insert houses TLD and radiochromic film. Institutions imaged, planned and treated the phantom. Plans were designed to deliver 30 Gy to the center of the target that was covered by an isodose line typically between 50% and 85%. The treated volume, assumed to be an ellipsoid defined by the prescription dose, was calculated from film profiles and compared to distances reported by the institution. Results: Between 2000 and 2010, 125 of the irradiations were performed with Gamma Knife and 509 were performed with linear accelerators. Most of these irradiations were performed as a method of independent verification of treatment delivery. Gamma Knife TLD and treatment volume results were on average 0.98±0.03 and 1.09±0.11, respectively. Linear accelerator TLD and treatment volume results were on average 1.00±0.03 and 1.04±0.15, respectively. The percentage of institutions, using accelerators, meeting the QA guidelines for dose to the target, treated volume, ratio of measured treated volume to target volume and minimum dose to target were 93%, 90%, 80%, and 80%, respectively. For Gamma Knifes the corresponding percentages were 91%, 98%, 88%, and 49%. The percent of institutions meeting all four criteria were 54% and 39% for accelerators and Gamma Knife, respectively. Conclusion: Though most SRS phantom irradiations met several of the individual criteria, only 50% met all four. Failure to plan adequate target coverage and/or deliver adequate target coverage contributed to the low percentages for minimum dose to target compliance. Work was supported by PHS grants CA10953 and CA81647 awarded by the NCI, DHHS.

Key concepts: Imaging phantom, Radiosurgery, Nuclear medicine, Thermoluminescent dosimeter, Dosimetry, Linear particle accelerator, Volume (thermodynamics), Radiation treatment planning