A Chronoamperometric Method To Estimate Ionophore Loss from Ion-Selective Electrode Membranes
Bradford D. Pendley, Ernö Lindner
Abstract
Bradford D. Pendley, Ernö Lindner
Abstract
A novel chronoamperometric method was developed to estimate the concentration of a neutral ionophore in fixed-site, dioctyl sebacate plasticized, poly(vinyl chloride)-based, ion-selective electrode membranes. The membranes contained between 0.5 and 16 mmol/kg valinomycin. The chronoamperometric technique was used to estimate the valinomycin concentration in freshly prepared membranes and after extraction of some of the ionophore from the membranes using dioctyl sebacate. Replicate measurements indicated a relative standard deviation in the calculated valinomycin concentration of less than 10%, and these values accurately represented the true concentration of valinomycin within 10%. The method permitted an estimate of the valinomycin concentration after valinomycin was leached from a membrane. The results of preliminary experiments using heparinized dog blood suggest that blood protein adsorption does not interfere qualitatively or quantitatively with the analysis.
OpenAlex reports 34 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A novel chronoamperometric method was developed to estimate the concentration of a neutral ionophore in fixed-site, dioctyl sebacate plasticized, poly(vinyl chloride)-based, ion-selective electrode membranes. The membranes contained between 0.5 and 16 mmol/kg valinomycin. The chronoamperometric technique was used to estimate the valinomycin concentration in freshly prepared membranes and after extraction of some of the ionophore from the membranes using dioctyl sebacate. Replicate measurements indicated a relative standard deviation in the calculated valinomycin concentration of less than 10%, and these values accurately represented the true concentration of valinomycin within 10%. The method permitted an estimate of the valinomycin concentration after valinomycin was leached from a membrane. The results of preliminary experiments using heparinized dog blood suggest that blood protein adsorption does not interfere qualitatively or quantitatively with the analysis.
Key concepts: Valinomycin, Ionophore, Chemistry, Membrane, Ion selective electrode, Chromatography, Vinyl chloride, Analytical Chemistry (journal)