Automated Enumeration of Immature Granulocytes

J. Fernandes, Yukio Hamaguchi

Open publisher page 61 citations

Abstract

The performance characteristics of the XE-2100 (Sysmex, Kobe, Japan) automated immature granulocyte (IG) count were studied. The automated IG count was compared with the manual morphology count and with a proposed reference flow cytometric count. The comparison data were analyzed by both least-squares and Passing-Bablok regression analysis. Long-term imprecision using preserved blood quality control specimens at different levels showed a range from 2.59% to 3.57% coefficient of variation (CV) for within-run imprecision and 3.57% to 6.85% CV for total imprecision. The within-run reproducibility performed using fresh blood on 3 different specimens showed a range from 5.55% to 8.24% CV. The counts were stable at both room temperature and after refrigeration for 24 hours.Passing-Bablok regression analysis showed excellent agreement between the proposed reference flow cytometric IG count and the XE-2100 IG count, while there was less agreement with the manual morphology count. Our results indicate that the automated IG count can replace the manual morphology count for IG counting in the clinical laboratory. The results also confirm that the flow cytometric IG count is superior to and can replace the manual morphology count as a reference method for IG counting.

About this research paper

What this paper is about

The performance characteristics of the XE-2100 (Sysmex, Kobe, Japan) automated immature granulocyte (IG) count were studied. The automated IG count was compared with the manual morphology count and with a proposed reference flow cytometric count. The comparison data were analyzed by both least-squares and Passing-Bablok regression analysis. Long-term imprecision using preserved blood quality control specimens at different levels showed a range from 2.59% to 3.57% coefficient of variation (CV) for within-run imprecision and 3.57% to 6.85% CV for total imprecision. The within-run reproducibility performed using fresh blood on 3 different specimens showed a range from 5.55% to 8.24% CV. The counts were stable at both room temperature and after refrigeration for 24 hours.Passing-Bablok regression analysis showed excellent agreement between the proposed reference flow cytometric IG count and the XE-2100 IG count, while there was less agreement with the manual morphology count. Our results indicate that the automated IG count can replace the manual morphology count for IG counting in the clinical laboratory. The results also confirm that the flow cytometric IG count is superior to and can replace the manual morphology count as a reference method for IG counting.

Why it matters

OpenAlex reports 61 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The performance characteristics of the XE-2100 (Sysmex, Kobe, Japan) automated immature granulocyte (IG) count were studied. The automated IG count was compared with the manual morphology count and with a proposed reference flow cytometric count. The comparison data were analyzed by both least-squares and Passing-Bablok regression analysis. Long-term imprecision using preserved blood quality control specimens at different levels showed a range from 2.59% to 3.57% coefficient of variation (CV) for within-run imprecision and 3.57% to 6.85% CV for total imprecision. The within-run reproducibility performed using fresh blood on 3 different specimens showed a range from 5.55% to 8.24% CV. The counts were stable at both room temperature and after refrigeration for 24 hours.Passing-Bablok regression analysis showed excellent agreement between the proposed reference flow cytometric IG count and the XE-2100 IG count, while there was less agreement with the manual morphology count. Our results indicate that the automated IG count can replace the manual morphology count for IG counting in the clinical laboratory. The results also confirm that the flow cytometric IG count is superior to and can replace the manual morphology count as a reference method for IG counting.

Key concepts: Coefficient of variation, Enumeration, Count data, Reproducibility, Absolute neutrophil count, Complete blood count, Nuclear medicine, Pathology