1988Lasers in Surgery and MedicineRequires access

Pulsed CO2 laser tissue ablation: Measurement of the ablation rate

Joseph T. Walsh, Thomas F. Deutsch

Open publisher page 191 citations

Abstract

Ablation of guinea pig skin using a CO2 laser emitting 2-mu sec-long pulses has been quantified by measuring the mass of tissue removed as a function of incident fluence per pulse. The mass-loss curves show three distinct regimes in which water evaporation, explosive tissue removal, and laser-induced plasma formation dominate. The data are fit to two models that predict that the mass removed depends either linearly or logarithmically on fluence. Although the data are best fit by a linear dependence upon fluence, plasma formation at high fluences prohibited obtaining data over a wide enough fluence range to differentiate unambiguously between the two models. Ablation efficiency, ablation thresholds, and the optical penetration depth at 10.6 micron were obtained from the measurements.

About this research paper

What this paper is about

Ablation of guinea pig skin using a CO2 laser emitting 2-mu sec-long pulses has been quantified by measuring the mass of tissue removed as a function of incident fluence per pulse. The mass-loss curves show three distinct regimes in which water evaporation, explosive tissue removal, and laser-induced plasma formation dominate. The data are fit to two models that predict that the mass removed depends either linearly or logarithmically on fluence. Although the data are best fit by a linear dependence upon fluence, plasma formation at high fluences prohibited obtaining data over a wide enough fluence range to differentiate unambiguously between the two models. Ablation efficiency, ablation thresholds, and the optical penetration depth at 10.6 micron were obtained from the measurements.

Why it matters

OpenAlex reports 191 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Ablation of guinea pig skin using a CO2 laser emitting 2-mu sec-long pulses has been quantified by measuring the mass of tissue removed as a function of incident fluence per pulse. The mass-loss curves show three distinct regimes in which water evaporation, explosive tissue removal, and laser-induced plasma formation dominate. The data are fit to two models that predict that the mass removed depends either linearly or logarithmically on fluence. Although the data are best fit by a linear dependence upon fluence, plasma formation at high fluences prohibited obtaining data over a wide enough fluence range to differentiate unambiguously between the two models. Ablation efficiency, ablation thresholds, and the optical penetration depth at 10.6 micron were obtained from the measurements.

Key concepts: Ablation, Fluence, Laser ablation, Laser, Penetration depth, Materials science, Plasma, Explosive material

Related papers

Back to paper searchBrowse research topicsOriginal source
Pulsed CO2 laser tissue ablation: Measurement of the ablation rate — Research Paper | ScholarLens