Quantitation of the large polypeptide glucagon by protein precipitation and LC/MS

David C. Delinsky, Kimberly T. Hill, Catherine A. White, Michael G. Bartlett

Open publisher page 31 citations

Abstract

We present a method for the quantitation of glucagon from rat plasma by protein precipitation and LC/MS. No internal standard was used, as a labeled standard was not available and similar peptides did not show comparable extraction characteristics to glucagon. The LC system included a Keystone C18, 300 A pore size column; a linear gradient was used with a mobile phase consisting of water and acetonitrile, each with 0.2% acetic acid and 0.02% trifluoroacetic acid. Glucagon was detected with the mass spectrometer in positive ion mode monitoring the 4+ charge state at m/z 871.7. The method had an approximated limit of detection of 1 ng/mL. The lower limit of quantitation (LLOQ) was 25 ng/mL (7.2 fmol/mL), which could be reduced with an appropriate internal standard. External calibration was used and calibration curves were found to be linear over the range from 25 to 1000 ng/mL (7.2 to 290 fmol/mL). The method showed a high degree of precision and accuracy both within and between runs at four validation points, including the LLOQ.

About this research paper

What this paper is about

We present a method for the quantitation of glucagon from rat plasma by protein precipitation and LC/MS. No internal standard was used, as a labeled standard was not available and similar peptides did not show comparable extraction characteristics to glucagon. The LC system included a Keystone C18, 300 A pore size column; a linear gradient was used with a mobile phase consisting of water and acetonitrile, each with 0.2% acetic acid and 0.02% trifluoroacetic acid. Glucagon was detected with the mass spectrometer in positive ion mode monitoring the 4+ charge state at m/z 871.7. The method had an approximated limit of detection of 1 ng/mL. The lower limit of quantitation (LLOQ) was 25 ng/mL (7.2 fmol/mL), which could be reduced with an appropriate internal standard. External calibration was used and calibration curves were found to be linear over the range from 25 to 1000 ng/mL (7.2 to 290 fmol/mL). The method showed a high degree of precision and accuracy both within and between runs at four validation points, including the LLOQ.

Why it matters

OpenAlex reports 31 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We present a method for the quantitation of glucagon from rat plasma by protein precipitation and LC/MS. No internal standard was used, as a labeled standard was not available and similar peptides did not show comparable extraction characteristics to glucagon. The LC system included a Keystone C18, 300 A pore size column; a linear gradient was used with a mobile phase consisting of water and acetonitrile, each with 0.2% acetic acid and 0.02% trifluoroacetic acid. Glucagon was detected with the mass spectrometer in positive ion mode monitoring the 4+ charge state at m/z 871.7. The method had an approximated limit of detection of 1 ng/mL. The lower limit of quantitation (LLOQ) was 25 ng/mL (7.2 fmol/mL), which could be reduced with an appropriate internal standard. External calibration was used and calibration curves were found to be linear over the range from 25 to 1000 ng/mL (7.2 to 290 fmol/mL). The method showed a high degree of precision and accuracy both within and between runs at four validation points, including the LLOQ.

Key concepts: Chemistry, Protein precipitation, Chromatography, Calibration curve, Trifluoroacetic acid, Detection limit, Mass spectrometry, Extraction (chemistry)

Related papers

Back to paper searchBrowse research topicsOriginal source
Quantitation of the large polypeptide glucagon by protein precipitation and LC/MS — Research Paper | ScholarLens