In-flight validation and calibration of the spectral and radiometric characteristics of the Airborne Visible/Infrared Imaging Spectrometer

Robert O. Green, J. E. Conel, J. S. Margolis, V. Carrère, Carol J. Bruegge, M. Rast, Gordon Hoover

Open publisher page 25 citations

Abstract

Through an in-flight calibration experiment at Rogers Dry Lake, California on September 20, 1989, the radiometric and spectral properties of AVIRIS were determined. In-flight spectral channel positions and the spectral response function in 10 regions of the AVIRIS spectral range, taking in all four spectrometers, are shown to agree closely with the corresponding parameters measured in the laboratory. The intraflight stability for the Rogers Dry Lake calibration site is better than 2 percent with the exception of the strong atmospheric water absorptions where the measured radiance is close to zero. This experiment has provided both direct generation of an in-flight spectral and radiometric calibration and validation of the laboratory calibration at the reported level accuracy.

About this research paper

What this paper is about

Through an in-flight calibration experiment at Rogers Dry Lake, California on September 20, 1989, the radiometric and spectral properties of AVIRIS were determined. In-flight spectral channel positions and the spectral response function in 10 regions of the AVIRIS spectral range, taking in all four spectrometers, are shown to agree closely with the corresponding parameters measured in the laboratory. The intraflight stability for the Rogers Dry Lake calibration site is better than 2 percent with the exception of the strong atmospheric water absorptions where the measured radiance is close to zero. This experiment has provided both direct generation of an in-flight spectral and radiometric calibration and validation of the laboratory calibration at the reported level accuracy.

Why it matters

OpenAlex reports 25 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Through an in-flight calibration experiment at Rogers Dry Lake, California on September 20, 1989, the radiometric and spectral properties of AVIRIS were determined. In-flight spectral channel positions and the spectral response function in 10 regions of the AVIRIS spectral range, taking in all four spectrometers, are shown to agree closely with the corresponding parameters measured in the laboratory. The intraflight stability for the Rogers Dry Lake calibration site is better than 2 percent with the exception of the strong atmospheric water absorptions where the measured radiance is close to zero. This experiment has provided both direct generation of an in-flight spectral and radiometric calibration and validation of the laboratory calibration at the reported level accuracy.

Key concepts: Radiance, Remote sensing, Radiometric calibration, Spectrometer, Calibration, Imaging spectrometer, Radiometric dating, Radiometry