SU-E-I-188: Potential for Dose Reduction in An Indirect Flat Panel Detector System Employing Irradiation Side Sampling

Nobukazu Tanaka, Tsutomu Akasaka, Hidetake Yabuuchi, Masafumi Ohki, Junji Morishita

Open publisher page 0 citations

Abstract

Purpose: To investigate possibility of dose reduction for a newly developed indirect flat panel detector (FPD) system employing irradiation side sampling (ISS), physical imaging properties were measured and compared with those of a conventional computed radiography (CR) system. Method and Materials:The FPD system employing ISS (CALNEO MT, Fuji Film) and the CR system (FCR PROFECT CS, Fuji Film) including the imaging plate (ST-VI, Fuji Film) were used in this study. The Wiener spectrum (WS), the presampled modulation transfer function (MTF), and the detective quantum efficiency (DQE) for both systems were measured. To evaluate image quality of the detector for chest radiography, the RQA 9 in the International Electrotechnical Commission (IEC) 62220-1 was used. Results: The WS of the FPD system showed considerably lower than that of the CR system at every spatial frequency. The WS of the FPD system with a 50 % dose reduction of the CR system was almost identical to that of the CR system. The presampled MTF of the FPD system was comparable to that of the CR system. The DQE of the FPD system obtained at zero spatial frequency was almost twice than that of the CR system. Conclusion: The radiographic noise property and the detection efficiency of the new ISS FPD were significantly superior to those of the conventional CR system. It is expected that the use of the new ISS FPD system for chest radiography can reduce the patient skin dose by 50 % compared with the conventional CR system.

About this research paper

What this paper is about

Purpose: To investigate possibility of dose reduction for a newly developed indirect flat panel detector (FPD) system employing irradiation side sampling (ISS), physical imaging properties were measured and compared with those of a conventional computed radiography (CR) system. Method and Materials:The FPD system employing ISS (CALNEO MT, Fuji Film) and the CR system (FCR PROFECT CS, Fuji Film) including the imaging plate (ST-VI, Fuji Film) were used in this study. The Wiener spectrum (WS), the presampled modulation transfer function (MTF), and the detective quantum efficiency (DQE) for both systems were measured. To evaluate image quality of the detector for chest radiography, the RQA 9 in the International Electrotechnical Commission (IEC) 62220-1 was used. Results: The WS of the FPD system showed considerably lower than that of the CR system at every spatial frequency. The WS of the FPD system with a 50 % dose reduction of the CR system was almost identical to that of the CR system. The presampled MTF of the FPD system was comparable to that of the CR system. The DQE of the FPD system obtained at zero spatial frequency was almost twice than that of the CR system. Conclusion: The radiographic noise property and the detection efficiency of the new ISS FPD were significantly superior to those of the conventional CR system. It is expected that the use of the new ISS FPD system for chest radiography can reduce the patient skin dose by 50 % compared with the conventional CR system.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: To investigate possibility of dose reduction for a newly developed indirect flat panel detector (FPD) system employing irradiation side sampling (ISS), physical imaging properties were measured and compared with those of a conventional computed radiography (CR) system. Method and Materials:The FPD system employing ISS (CALNEO MT, Fuji Film) and the CR system (FCR PROFECT CS, Fuji Film) including the imaging plate (ST-VI, Fuji Film) were used in this study. The Wiener spectrum (WS), the presampled modulation transfer function (MTF), and the detective quantum efficiency (DQE) for both systems were measured. To evaluate image quality of the detector for chest radiography, the RQA 9 in the International Electrotechnical Commission (IEC) 62220-1 was used. Results: The WS of the FPD system showed considerably lower than that of the CR system at every spatial frequency. The WS of the FPD system with a 50 % dose reduction of the CR system was almost identical to that of the CR system. The presampled MTF of the FPD system was comparable to that of the CR system. The DQE of the FPD system obtained at zero spatial frequency was almost twice than that of the CR system. Conclusion: The radiographic noise property and the detection efficiency of the new ISS FPD were significantly superior to those of the conventional CR system. It is expected that the use of the new ISS FPD system for chest radiography can reduce the patient skin dose by 50 % compared with the conventional CR system.

Key concepts: Detective quantum efficiency, Flat panel detector, Computed radiography, Optical transfer function, Detector, Digital radiography, Radiography, Optics

Related papers

Back to paper searchBrowse research topicsOriginal source
SU-E-I-188: Potential for Dose Reduction in An Indirect Flat Panel Detector System Employing Irradiation Side Sampling — Research Paper | ScholarLens