Implementation of Electron-Transfer Dissociation on a Hybrid Linear Ion Trap−Orbitrap Mass Spectrometer

Graeme C. McAlister, Doug Phanstiel, David M. Good, W. Travis Berggren, Joshua J. Coon

Open full text 172 citations

Abstract

We describe the adaptation of a hybrid quadrupole linear ion trap-orbitrap mass spectrometer to accommodate electron-transfer ion/ion reactions (ETD) for peptide and protein characterization. The method utilizes pulsed, dual electrospray ion sources and requires minimal instrument modification. Switching between cation and reagent anion injection schemes is automated and accomplished within a few hundred milliseconds. Ion/ion reactions are conducted within the linear ion trap, after which the c- and z-type product ions are passed to the orbitrap for high-resolution m/z analysis. With this arrangement, mass accuracies are typically measured to within 2 ppm at a resolving power of approximately 60 000. Using large peptides and intact proteins, we demonstrate such capabilities will accelerate our ability to interrogate high-mass species. To illustrate compatibility with automated data-dependent analysis and subsequent data processing, we couple the technique with an online chromatographic separation of a yeast whole-cell lysate followed by peptide identification using ProSight PC. Fairly long pulsing times and relatively low ET efficiency, as compared to conventional ETD instrumentation, are the main drawbacks of this approach. Still, our results suggest that the implementation of ETD on sensitive, high-resolution, and high-mass accuracy hybrid instrumentation, such as the orbitrap, will substantially propel the emergent fields of middle- and top-down proteomics.

About this research paper

What this paper is about

We describe the adaptation of a hybrid quadrupole linear ion trap-orbitrap mass spectrometer to accommodate electron-transfer ion/ion reactions (ETD) for peptide and protein characterization. The method utilizes pulsed, dual electrospray ion sources and requires minimal instrument modification. Switching between cation and reagent anion injection schemes is automated and accomplished within a few hundred milliseconds. Ion/ion reactions are conducted within the linear ion trap, after which the c- and z-type product ions are passed to the orbitrap for high-resolution m/z analysis. With this arrangement, mass accuracies are typically measured to within 2 ppm at a resolving power of approximately 60 000. Using large peptides and intact proteins, we demonstrate such capabilities will accelerate our ability to interrogate high-mass species. To illustrate compatibility with automated data-dependent analysis and subsequent data processing, we couple the technique with an online chromatographic separation of a yeast whole-cell lysate followed by peptide identification using ProSight PC. Fairly long pulsing times and relatively low ET efficiency, as compared to conventional ETD instrumentation, are the main drawbacks of this approach. Still, our results suggest that the implementation of ETD on sensitive, high-resolution, and high-mass accuracy hybrid instrumentation, such as the orbitrap, will substantially propel the emergent fields of middle- and top-down proteomics.

Why it matters

OpenAlex reports 172 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We describe the adaptation of a hybrid quadrupole linear ion trap-orbitrap mass spectrometer to accommodate electron-transfer ion/ion reactions (ETD) for peptide and protein characterization. The method utilizes pulsed, dual electrospray ion sources and requires minimal instrument modification. Switching between cation and reagent anion injection schemes is automated and accomplished within a few hundred milliseconds. Ion/ion reactions are conducted within the linear ion trap, after which the c- and z-type product ions are passed to the orbitrap for high-resolution m/z analysis. With this arrangement, mass accuracies are typically measured to within 2 ppm at a resolving power of approximately 60 000. Using large peptides and intact proteins, we demonstrate such capabilities will accelerate our ability to interrogate high-mass species. To illustrate compatibility with automated data-dependent analysis and subsequent data processing, we couple the technique with an online chromatographic separation of a yeast whole-cell lysate followed by peptide identification using ProSight PC. Fairly long pulsing times and relatively low ET efficiency, as compared to conventional ETD instrumentation, are the main drawbacks of this approach. Still, our results suggest that the implementation of ETD on sensitive, high-resolution, and high-mass accuracy hybrid instrumentation, such as the orbitrap, will substantially propel the emergent fields of middle- and top-down proteomics.

Key concepts: Chemistry, Orbitrap, Electron-transfer dissociation, Quadrupole ion trap, Mass spectrometry, Hybrid mass spectrometer, Ion trap, Ion