2000The Journal of UrologyRequires access

MANAGEMENT OF CUFF EROSION OF THE DOUBLE CUFF ARTIFICIAL URINARY SPHINCTER

BRADLEY B. BELL, John J. Mulcahy

Open publisher page 10 citations

Abstract

PURPOSE: We report on a series of patients with erosion of 1 cuff of an AMS800 double cuff artificial urinary sphincter and determine the success rate of removing the eroded cuff, leaving the other components behind and converting the device to a single cuff system. MATERIALS AND METHODS: In 9 patients with a double cuff artificial sphincter 1 eroded cuff was removed and the system was converted to a single cuff sphincter. RESULTS: Followup was based on an incontinence scoring system. In 1 patient the remaining cuff was removed because it also eroded. In another patient a wound infection developed 5 months after placement and the entire device was removed. The remaining 7 patients had better continence scores than before artificial sphincter placement but each remarked that continence was better with 2 cuffs. CONCLUSIONS: An eroded cuff of a double cuff artificial urinary sphincter system can be successfully removed and the device can be converted to a single cuff system.

About this research paper

What this paper is about

PURPOSE: We report on a series of patients with erosion of 1 cuff of an AMS800 double cuff artificial urinary sphincter and determine the success rate of removing the eroded cuff, leaving the other components behind and converting the device to a single cuff system. MATERIALS AND METHODS: In 9 patients with a double cuff artificial sphincter 1 eroded cuff was removed and the system was converted to a single cuff sphincter. RESULTS: Followup was based on an incontinence scoring system. In 1 patient the remaining cuff was removed because it also eroded. In another patient a wound infection developed 5 months after placement and the entire device was removed. The remaining 7 patients had better continence scores than before artificial sphincter placement but each remarked that continence was better with 2 cuffs. CONCLUSIONS: An eroded cuff of a double cuff artificial urinary sphincter system can be successfully removed and the device can be converted to a single cuff system.

Why it matters

OpenAlex reports 10 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: We report on a series of patients with erosion of 1 cuff of an AMS800 double cuff artificial urinary sphincter and determine the success rate of removing the eroded cuff, leaving the other components behind and converting the device to a single cuff system. MATERIALS AND METHODS: In 9 patients with a double cuff artificial sphincter 1 eroded cuff was removed and the system was converted to a single cuff sphincter. RESULTS: Followup was based on an incontinence scoring system. In 1 patient the remaining cuff was removed because it also eroded. In another patient a wound infection developed 5 months after placement and the entire device was removed. The remaining 7 patients had better continence scores than before artificial sphincter placement but each remarked that continence was better with 2 cuffs. CONCLUSIONS: An eroded cuff of a double cuff artificial urinary sphincter system can be successfully removed and the device can be converted to a single cuff system.

Key concepts: Cuff, Artificial urinary sphincter, Medicine, Surgery, Sphincter, Urinary incontinence, Urethral sphincter, Urinary system