An impedimetric approach for accurate particle sizing using a microfluidic Coulter counter

Ashish V. Jagtiani, Joan Carletta, Jiang Zhe

Open publisher page 26 citations

Abstract

In this paper, we present the design, impedimetric characterization and testing of a microfabricated Coulter counter for particle size measurement that uses a pair of thin film coplanar Au/Ti electrodes. An electrical equivalent circuit model of the designed device is analyzed. Accurate measurement of particle size was achieved by operating the device at a frequency for which the overall impedance is dominated by the channel resistance. A combination of design features, including the use of a pair of sensing electrodes with a surface area of 100 µm by 435 µm, a spacing of 1785 µm between the two sensing electrodes and a 350 µm long microchannel, ensures that this resistance dominates over a range of relatively low frequencies. The device was characterized for NaCl electrolyte solutions with different ionic concentrations ranging from 10−5 to 0.1 M. Results proved that the resistive behavior of the sensor occurs over a range of relatively low frequencies for all tested concentrations. The Coulter counter was then used to detect 30 µm polystyrene particles at a selected excitation frequency. Testing results demonstrated that the device can accurately measure particle sizes with small error. The design can be extended to ac Coulter counters with sub-micron sensing channels. Analysis of three designs of ac Coulter counters including sub-micron sensing channels using the electrical equivalent circuit model predicts that they can be operated at even lower frequencies, to accurately size nanoscale particles.

About this research paper

What this paper is about

In this paper, we present the design, impedimetric characterization and testing of a microfabricated Coulter counter for particle size measurement that uses a pair of thin film coplanar Au/Ti electrodes. An electrical equivalent circuit model of the designed device is analyzed. Accurate measurement of particle size was achieved by operating the device at a frequency for which the overall impedance is dominated by the channel resistance. A combination of design features, including the use of a pair of sensing electrodes with a surface area of 100 µm by 435 µm, a spacing of 1785 µm between the two sensing electrodes and a 350 µm long microchannel, ensures that this resistance dominates over a range of relatively low frequencies. The device was characterized for NaCl electrolyte solutions with different ionic concentrations ranging from 10−5 to 0.1 M. Results proved that the resistive behavior of the sensor occurs over a range of relatively low frequencies for all tested concentrations. The Coulter counter was then used to detect 30 µm polystyrene particles at a selected excitation frequency. Testing results demonstrated that the device can accurately measure particle sizes with small error. The design can be extended to ac Coulter counters with sub-micron sensing channels. Analysis of three designs of ac Coulter counters including sub-micron sensing channels using the electrical equivalent circuit model predicts that they can be operated at even lower frequencies, to accurately size nanoscale particles.

Why it matters

OpenAlex reports 26 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

In this paper, we present the design, impedimetric characterization and testing of a microfabricated Coulter counter for particle size measurement that uses a pair of thin film coplanar Au/Ti electrodes. An electrical equivalent circuit model of the designed device is analyzed. Accurate measurement of particle size was achieved by operating the device at a frequency for which the overall impedance is dominated by the channel resistance. A combination of design features, including the use of a pair of sensing electrodes with a surface area of 100 µm by 435 µm, a spacing of 1785 µm between the two sensing electrodes and a 350 µm long microchannel, ensures that this resistance dominates over a range of relatively low frequencies. The device was characterized for NaCl electrolyte solutions with different ionic concentrations ranging from 10−5 to 0.1 M. Results proved that the resistive behavior of the sensor occurs over a range of relatively low frequencies for all tested concentrations. The Coulter counter was then used to detect 30 µm polystyrene particles at a selected excitation frequency. Testing results demonstrated that the device can accurately measure particle sizes with small error. The design can be extended to ac Coulter counters with sub-micron sensing channels. Analysis of three designs of ac Coulter counters including sub-micron sensing channels using the electrical equivalent circuit model predicts that they can be operated at even lower frequencies, to accurately size nanoscale particles.

Key concepts: Coulter counter, Sizing, Microfluidics, Particle (ecology), Materials science, Nanotechnology, Chemistry, Biology