Separation of an arterial catheter from its hub

G. Knottenbelt, J. Van Der Westhuisen

Open publisher page 1 citations

Abstract

We would like to report an incident where the catheter of a Vygon Leadercath 20G arterial catheter (Vygon UK, Cirencester, UK) separated from its connection hub and lodged in the patient's radial artery. The Leadercath cannula uses the Seldinger technique in which the arterial catheter is introduced over a guidewire placed through a needle. In replacing arterial lines, it is common practice to introduce a Seldinger wire through the catheter already in place, then remove the original arterial catheter, and introduce the new arterial line over the wire. In this incident the introduction of the Seldinger wire dislodged the body of the catheter from the connection hub, leaving the catheter in the patient's radial artery (Fig. 4). The wire was from an identical, replacement arterial catheter. No excessive resistance to the introduction of the wire was felt and the soft end of the wire was used. Following discussion with the vascular surgeons we agreed to treat this incident expectantly as patient's circulation was not compromised, and the patient had a poor prognosis. The patient subsequently died from unrelated problems. Vygon arterial catheter separated from the hub (top) and an intact catheter (below). We have been informed by Vygon UK that to their knowledge this complication has not occurred before. It is clear that care should be taken when replacing arterial lines in this way.

About this research paper

What this paper is about

We would like to report an incident where the catheter of a Vygon Leadercath 20G arterial catheter (Vygon UK, Cirencester, UK) separated from its connection hub and lodged in the patient's radial artery. The Leadercath cannula uses the Seldinger technique in which the arterial catheter is introduced over a guidewire placed through a needle. In replacing arterial lines, it is common practice to introduce a Seldinger wire through the catheter already in place, then remove the original arterial catheter, and introduce the new arterial line over the wire. In this incident the introduction of the Seldinger wire dislodged the body of the catheter from the connection hub, leaving the catheter in the patient's radial artery (Fig. 4). The wire was from an identical, replacement arterial catheter. No excessive resistance to the introduction of the wire was felt and the soft end of the wire was used. Following discussion with the vascular surgeons we agreed to treat this incident expectantly as patient's circulation was not compromised, and the patient had a poor prognosis. The patient subsequently died from unrelated problems. Vygon arterial catheter separated from the hub (top) and an intact catheter (below). We have been informed by Vygon UK that to their knowledge this complication has not occurred before. It is clear that care should be taken when replacing arterial lines in this way.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We would like to report an incident where the catheter of a Vygon Leadercath 20G arterial catheter (Vygon UK, Cirencester, UK) separated from its connection hub and lodged in the patient's radial artery. The Leadercath cannula uses the Seldinger technique in which the arterial catheter is introduced over a guidewire placed through a needle. In replacing arterial lines, it is common practice to introduce a Seldinger wire through the catheter already in place, then remove the original arterial catheter, and introduce the new arterial line over the wire. In this incident the introduction of the Seldinger wire dislodged the body of the catheter from the connection hub, leaving the catheter in the patient's radial artery (Fig. 4). The wire was from an identical, replacement arterial catheter. No excessive resistance to the introduction of the wire was felt and the soft end of the wire was used. Following discussion with the vascular surgeons we agreed to treat this incident expectantly as patient's circulation was not compromised, and the patient had a poor prognosis. The patient subsequently died from unrelated problems. Vygon arterial catheter separated from the hub (top) and an intact catheter (below). We have been informed by Vygon UK that to their knowledge this complication has not occurred before. It is clear that care should be taken when replacing arterial lines in this way.

Key concepts: Medicine, Catheter, Seldinger technique, Arterial line, Arterial catheter, Cannula, Radial artery, Surgery