Repeatability of successive measurements with a portable nitric oxide analyser in patients with suggested or diagnosed asthma
Reijo Takalo, Päivi Piirilä, Anssi Sovijärvi
Abstract
Reijo Takalo, Päivi Piirilä, Anssi Sovijärvi
Abstract
The objective of our study was to determine the repeatability of successive measurements of exhaled nitric oxide (NO) with a portable device (NIOX MINO). We assembled a sample of 332 pairs of repeated NO measurements acquired in the same session. The mean of the two successive measurements varied from 4.0 to 279.5 parts per billion (ppb). The mean coefficient of repeatability (CoR) of the entire study sample was 2.4 ppb. The difference between paired measurements increased slightly with increasing fractional exhaled nitric oxide (FENO); for the mean FENO of 24.0 ppb the calculated CoR was 2.2 and for the FENO of 54 ppb it was 3.2 ppb. The difference between paired observations for FENO values <50 ppb was significantly lower than that for values >50 ppb (p<0.001). The demonstrated repeatability of successive measurements with the device supports the view that, when the exhalation manoeuvre is technically valid and FENO is within normal limits, repeated determinations of FENO are generally not necessary for screening. In other cases, the mean value of two successive measurements would represent a more reliable measurement result.
OpenAlex reports 14 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The objective of our study was to determine the repeatability of successive measurements of exhaled nitric oxide (NO) with a portable device (NIOX MINO). We assembled a sample of 332 pairs of repeated NO measurements acquired in the same session. The mean of the two successive measurements varied from 4.0 to 279.5 parts per billion (ppb). The mean coefficient of repeatability (CoR) of the entire study sample was 2.4 ppb. The difference between paired measurements increased slightly with increasing fractional exhaled nitric oxide (FENO); for the mean FENO of 24.0 ppb the calculated CoR was 2.2 and for the FENO of 54 ppb it was 3.2 ppb. The difference between paired observations for FENO values <50 ppb was significantly lower than that for values >50 ppb (p<0.001). The demonstrated repeatability of successive measurements with the device supports the view that, when the exhalation manoeuvre is technically valid and FENO is within normal limits, repeated determinations of FENO are generally not necessary for screening. In other cases, the mean value of two successive measurements would represent a more reliable measurement result.
Key concepts: Repeatability, Exhaled nitric oxide, Exhalation, Coefficient of variation, Analyser, Limits of agreement, Medicine, Chemistry