Determination of Tetracycline and Related Compounds by High-Performance Liquid Chromatography
J. Y. C. Hon, L. R. Murray
Abstract
J. Y. C. Hon, L. R. Murray
Abstract
An isocratic high-performance liquid chromatography method for the determination of tetracycline and its related compounds is described. The method uses a reverse phase (C18) column, a modified acetonitrile/water mobile phase, and banzoic acid as the internal standard. Elution of all compounds of interest is complete within seven minutes. Results are presented for thirteen commercial capsule formulations and are compared with results by microbiological assay and thin-layer chromatographic methods.
OpenAlex reports 18 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
An isocratic high-performance liquid chromatography method for the determination of tetracycline and its related compounds is described. The method uses a reverse phase (C18) column, a modified acetonitrile/water mobile phase, and banzoic acid as the internal standard. Elution of all compounds of interest is complete within seven minutes. Results are presented for thirteen commercial capsule formulations and are compared with results by microbiological assay and thin-layer chromatographic methods.
Key concepts: Tetracycline, Chromatography, Chemistry, Antibiotics, Biochemistry